Instrument Database:
ATAS GL International B.V. - MP711 Micro-flow pump
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
ATAS GL International B.V.
|
Categories |
Chromatography: LC-Pump Chromatography: LC-Pump: Capillary |
Highly accurate and reproducible micro-flows for capillary Nano LC and LCMS
Solvent-delivery without any flow splitter Pulse-free solvent-delivery
The micro-flow pump has two independent highly precise drive systems to generate a continuous and constant flow with flow accuracy of ± 2% and flow precision of 0.3%.
|
Specifications |
|
Flow-rate range: Isocratic mode: | 0.01 μl/min to 50 μl/min.
| High-pressure gradient mode: | 1.0 μl/min to 10 μl/min.
| Flow-rate increment | 0.01-1.00 μl/min: 0.01 μl/min increment. 1-50 μl/min: 0.1 μl/min increment. | Flow-rate Accuracy: | ±2 % (distilled water, 20ºC, 5.0ul/min, 5-34.0 MPa). | Maximum Operating Pressure: | 34.0 Mpa. | Accuracy of the Displayed Pressure: | ±2% or ±0.4 MPa, whichever is larger. | Pressure Limit: | Upper and Lower limits (1.0 to 34 MPa) | Purge/Drain Valve: | Automatic open/close, rapid purge at a flow-rate of 1.0 ml/min. | Wetted Materials: | 316 SS, PEEK, PTFE, ruby, sapphire | Program: | Time and Gradient programs; all programs have 30 steps and 10 files. | Input signals: | pump on/off, program start/stop, stop/start from external device. | Output signal: | ERROR, READY, EVENT1, and EVENT2. | Communication: | RS232C, and special network communication. | Power Requirements: | Full range for input voltage of 100-240 V AC, 50/60 Hz. | Dimensions: | W260xH203xD458. | Weight: | 15 kg. | Operating Temperature Range: | 5 to 35°C. | Operating Humidity: | 35 to 85%. | Power Consumption: | Max. 120 VA. |
|
|