Instrument Database:
Knauer GmbH - Smartline UV Detector 2600
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
Knauer GmbH
|
Categories |
Chromatography: LC-Detector: UV/Vis |
Multi-wavelength spectro-photometer based on diode array technology.
The Smartline UV Detector 2600 can simultaneously measure at up to four freely programmable detection channels with high sensitivity and continuously record spectra in a selectable scan-range. The signals of the four channels are additionally distributed on the four analog outputs. With the level and event outputs, it is even possible to set up software-independent solvent recycling by just adding a 3-port solenoid valve. Together with KNAUER ChromGate® or EuroChrom® software many functions that are typical of a diode array detector are supported including multi-wavelength operation, 2-D and 3-D chromatograms, peak purity, spectra libraries, and more.
|
Specifications |
|
Diode Array 256 diodes (1.25 nm /diode)
Wavelength Range 190 - 500 nm
Light Source deuterium
Channels 4 (data recording)
Wavelength Accuracy / Bandwidth ≤ 1 nm (254 nm) / 4-25 nm (selectable)
Spectra Scan max. 10 scans per second
Noise / Drift (ASTM E1657-94) ≤ 1x10-5 AU / ≤ 5x10-4 AU/h
Analog Output (Integrator) 4 x ± 1 V (scalable), 20-bit
Control RS-232, remote control strip, touchpad
Miscellaneous Programming: max. 20 programs, detailed GLP report, automatic wavelength verification via internal holmium oxide filter, LC display: 2 lines, 24 characters each, universal power supply: 90-260 V, 47-63 Hz
Dimensions / Weight 226 x 135 x 390 (W x H x D) / 5.7 kg
|
|