Instrument Database:
Shimadzu Corporation - TMA-50/50H
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
Shimadzu Corporation
|
Categories |
Thermal Analysis: TMA |
Thermomechanical Analyzer In addition to thermal expansion, the TMA-50/50H is also capable of S-S measurement of fiber and film. Because load control is performed by means of a force coil, sample setting is easy. Glass and ceramics can be measured by switching the measurement method to the differential expansion method.
- Temperature range: Room temperature to 1000/1500°C (-140°C option)
- Measurement range: Displacement ±250 µm, ±2500 µm
- Load ±100 g, 500 g Sample dimensions: 8 ø x 20 mm, 5 (width) x 1 (thickness) x 20 mm (length
Unmatched Flexibility In Material Characterization Variety of Measurement Modes Wide Dynamic Measuring Range Differential Expansion Measurements Extensive Load Selection Customized Programming
|
Specifications |
|
| TMA-50 | TMA-50H | Temperature Range | Ambient to 1000°C | Ambient to 1500°C | Measuring range | displacement: 2500 m , 250 m load : 500g, 100g | Sample dimensions | 8X20mm or less 5X20mm or less (for differential measurement) | Sample load | 500g max (10mg dial in) | Film Samples | 5 (width) X 1 (thickness) X 20 (length)mm or less | Hold time | 0 to 999min or 0 to 999hour | Temperature program | 99 steps maximum | Temperature program file | up to 100files | Measurement mode | Expansion, Penetration, Elongation | Differential Expansion | Load mode | - Constant rate load: up to 50 steps
- Constant rate elongation: up to 50 steps
- Shrink stress (Isostatic Load): max. 100g, max. 500g
- Cyclic load: frequency 0.001 to 1Hz
| Cooling system | Air cooled | Signal output | Analog and digital | Atmosphere | Air, inert gas | Dimensions | W173xD550xH795mm | Power supply | AC100V, 120V, 220V, 240V, 500VA, 50/60Hz | AC100V, 120V, 220V, 240V, 1.2kVA, 50/60Hz | |
|
|