Instrument Database:
Gilson, Inc - 925 PC Workstation
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
Gilson, Inc
|
Categories |
Automation: Workstation |
The 925 PC Workstation automates the preparation of high-throughput protein crystallography. It is engineered to meet the demand for high-throughput reagent screening, yet can provide large, high-quality crystals suitable for X-ray diffraction.
- Processes up to four sitting drop plates per batch
- Significantly reduces the time and expense needed to determine strategies for optimization experiments—processes 96 samples in less than 7 minutes
- Flexibility of using time-tested hanging or sitting drop vapor diffusion techniques—no need to validate new methods
- Multitool head allows for plate gripper and 96- and eight-channel pipettors for fast transfer of screening and protein solutions
- Eliminates carryover and cross-contamination by using disposable pipette tips for accurate, repeatable and dependable results (0.5 µL of 30% PEG 8000 at <3% CV)
- Deck layout holds plates, tips, screening solutions and protein reservoirs on the same bed, allowing for complete and quick automation
- Aspirates protein in volumes as low as 0.5 µL, which saves precious protein sample and costly reagents
Applications
- High-throughput crystallography rapid screening protocols
- Rapid and efficient screening of large numbers of different proteins
|
Specifications |
|
Liquid Handling Reproducibility: <3% CV with 0.5 µL of 30% PEG 8000 Accuracy: 101% with 0.5 µL of 30% PEG 8000
Arm Speed (max.) X/Y: 152.4 cm/sec (60 in./sec) Z: 25.4 cm/sec (10 in./sec)
Compressed Air Requirements 80 psi (5.3 bar) @ 2 cfm
Environmental Conditions Indoor use Altitude: up to 2000 m Temperature range: 5–40°C Air pressure: 75–105 kPa Pollution degree: 1 or 2, in accordance with IEC 66 Humidity: Maximum relative humidity 80% for temperatures up to 31°C, decreasing linearly to 50% relative humidity at 40°C
Maximum Syringe Flow Rates for Water 1000 µL/sec; adjustable from 75 to 1000 µL/sec
Manufacturing Standards Meets applicable Safety and EMC certification standards; UL and CE certified
Power Requirements Frequency: 50–60 Hz Voltage: 100–120 or 220–240V; mains voltage fluctuations not to exceed ±10% of the nominal voltage Current rating: 10A for 115V AC (20A with chiller) or 7A for 230V AC (15A with chiller)
Probe Positioning Performance Accuracy: ±0.2 mm (0.008 in.) in X/Y/Z dimensions Repeatability: ±0.1 mm (0.004 in.) in X/Y/Z dimensions
Sampler Type Gantry style X/Y/Z with stationary rack design
Software Via Gilson 745 Control Software
Syringe Capacity (maximum) 96-channel: 200 µL 8-channel: 50 µL Single-channel: 200 µL
Syringe Pump Internal, high-precision, piston
Temperature Range 5–40°C* *Actual temperature is a function of the ambient temperature and the style of microplates used.
Typical Number of Experimental Drops per Hour 1110+
Typical Volume of Protein Required 100 µL (about 10 µL per 8 sample vials)
Typical Volume of Screening Solution Required 50 µL per 96 wells
Vacuum Requirements 25” Hg @ 2 cfm
Vertical Punch Strength 96-Well: 4.5 kg (10 lbs.) Gripper: 8.2 kg (18 lbs.)
Dimensions* (w x d x h) *Include air gap and connectors 122 x 96 x 88 cm (48 x 37.8 x 34.6 in.)
Weight 132 kg (291 lbs.)
|
|