Instrument Database:
Beijing Titan Instruments Co., Ltd - AFS-930
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
Beijing Titan Instruments Co., Ltd
|
Categories |
Spectrometer ( Atom. ): AFS |
1 The instrument purpose Routing monitoring and analyses of As, Sb, Bi, Hg, Se, Te, Sn, Ge, Pb, Zn, Cd trace elements in total; 2 System configure requirement 2.1 Instrument type: Fully-automatic sequential pump injection hydride generation double –channel atomic fluorescence spectrometer 2.2 Configure content 2.2.1 Atomic fluorescence spectrometer 2.2.2 Automatic Injection Sampling System:>116units 2.2.3 Data process system
3 Working conditions 3.1 Environment temperature: 10 - 35°C 3.2 Relative humidity: 10% - 90% 3.3 Power supply: 220V (AC) 50Hz 3.4 Working power supply: 220v±10%50Hz
4、Technical requirements 4.1 Light source: Adopt special Hollow cathode core lamp tailor-made as light source controlled by system of Aggregate-pulse. 4.2 Optical system : Collection lens of short focus, non-chromatic dispersion system and the complete airtight system light avoidable and modulation light system 4.3 * Fully-automatic sequential pump injection hydride generation Device of pellicle separator for vapor dispelled 4.4 Advanced shield Quartz Atomizer, Atomizer argon --hydrogen flame 4.5 *automatically create standard curve, auto-dilution of high concentration samples 4.6 Retrenchment style and flux automatic controlled Gas route system; 4.7 With standard RS-232/RS-458 interface 4.8 Detection Limit (ug/l): Se、Te、Bi、As and so on<0.02μg/L Hg≤0.002μg/L Cd≤0.002μg/L 4.9 Related standard deviation: RSD≤1.0% 4.10 Dynamic range: three Q,TY degrees.
|
Specifications |
|
|
|