Instrument Database:
Beijing Rayleigh Analytical Instrument Corporation (BRAIC) - VIS-7220/7220G UV-9100/9200 Spectrophotometer
|
|
|
|
|
|
|
Year of introduction |
|
Status |
historical ( out of sale ) |
Company |
Beijing Rayleigh Analytical Instrument Corporation (BRAIC)
|
Categories |
Spectrometer ( Molec. ): UV/VIS |
- 2nm spectral bandwidth for accuracy.
- Controlled by microcomputer, LED display.
- Working curve with three points used for calculating concentrations.
- Parallel interface is available for printing.
- Easy operation, high flexibility.
- Compact, rugged construction and modern design.
|
Specifications |
|
Wavelength 330 - 800nm (VIS-7220) 320 - 1100nm (VIS-7220G) 200 - 800nm (UV-9100) 190 - 1000nm (UV-9200)
Wavelength accuracy ±2.0nm
Wavelength reproducibility 1.0nm
Photometric accuracy ±0.3%T
Photometric reproducibility 0.3%T
Stray light 0.6%T(NaNO2 at 360nm) for (VIS-7220/VIS7220G) 0.6%T (NaNO2 at 360nm,Nal at 220nm) for (UV-9100/9200)
Spectral bandwidth 2nm
0%T stability 0.2%T(3min)
100%T stability 0.5%T(3min)
Operation mode T: 0 - 110%; A: 0 - 2; C: Direct-reading concentration 0 - 99
Display mode LED display with four digits
Computer interface RS232C(optional)
Printer interface Universal parallel interface
Monochromator Single-beam, C-T type, grating 1200L/mm
Receiver Phototube (VIS-7220) Silicon Photodiode (VIS-7220G) Phototube (UV-9100) Silicon Photodiode (UV-9200)
Light source Tungsten halogen lamp (VIS-7220/VIS7220G) Tungsten halogen lamp, D2 lamp (UV-9100/9200)
Power Requirement 220/110V±10%,50/60HZ
Power consumption 80W (VIS-7220/VIS7220G) 120W (UV-9100/9200)
Dimensions 470×370×188mm
Net weight 10kg (VIS-7220/VIS7220G) 13kg (UV-9100/9200)
|
|