Instrument Database:
ECOM spol. s.r.o. - LCD 2073A Dual UV Detector
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
ECOM spol. s.r.o.
|
Categories |
Chromatography: LC-Detector: Diode Array |
LCD 2073A Dual UV Detector
is a new type of UV detector, which allows measuring absorbance of two wavelengths simultaneously in one cell. This unit is used in liquid chromatography to verify analyzed samples by means of a second wavelength or in situations when some peaks absorb on different wavelengths. It is possible to use the detector in analytical, preparative or flash applications by using different cells. Noise level at 254 nm is ± 1 x 10-4 AU with an empty cell.
The unit’s DAD (diode array detector) design offers many advantages:
- A wavelength setting may be changed in steps of 1nm
- The unit is easy for operating, reference and sample signal are available for detector diagnostics
- Output signal is it available in digital and analogue form on the output connector
- Work hours of lamp are counted using build in counter.
- Detector cell is easy replaceable from the side of the detector
- it is possible to control the unit using the external logic signals or an RS232 interface
This unit is an OEM aplication of TOPAZ Dual UV Detector
|
Specifications |
|
Wavelength range 200 - 320 nm Spectral half-width 8 nm Accuracy of adjustment ± 1 nm Reproducibility ± 0.5 nm Light source Deuterium discharge lamp IST WL 24 198 Noise level at empty cell (240 nm, TC 1 s) ± 1 x 10-4 AU Drift at empty cell (240 nm after 1 h) 5 x 10-4 AU/hr Materials in contact with mobile phase PTFE; fused silica, stainless steel, Vespel, PEEK Time constant 1 s Output for integrator 1 V/AU Interface RS 232 Power supply 230 V ± 10%, 50 Hz , 115/100 V± 10%, 50-60 Hz Power input 75 VA Dimensions (W x H x D) 200 x 178 x 398 mm Weight 12 kg
|
|