Instrument Database:
LAR Process Analysers AG - Elox100plus II COD Analyzer
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
LAR Process Analysers AG
|
Categories |
Dedicated Analyzer: Oxygen demand: COD |
The measuring point extension with stream switching and dual channel softwareThe Elox100plus II of LAR Process Analysers AG can deliver two new COD values every 4 minutes at maximum frequency of analysis. Additionally, the measurement frequency and desired range can be individually configured for both measurement points. The new Elox100plus can have this option retro-fitted.
|
Specifications |
|
Measurement and Sample Preparation
Type of measurement | COD measurement by electrochemical oxidation | Measurement Range | for every channel ranging from 1 - 100 mg/l COD to 1 - 100,000 mg/l COD | Actual Value | every 4 minutes per channel | Response Time | 30 seconds | Sample Preparation | maintenance-free particle separator |
Operation and Data Output
Graphic-LCD-screen, high resolution, back-lit | Autostart-function | Self-explanatory Software including maintenance checklists and support | Industry standard data interface | Data storage on 3.5" floppy disk |
Connections
Waste water, drain | tube 30 mm ID or threaded 32 mm OD or as specified | Electrical Power | 230 / 100 V ~, 50 / 60 Hz, 50 VA | Analog Output | 0 / 4 - 20 mA | Serial Interface | RS 232 for remote control malfunction alarm Life-Zero connection for printer |
Dimensions and Weight
Cabinet | Stainless Steel IP 64 (NEMA 4X) | Dimensions | 720 x 600 x 420 mm (H x W x D) 28.4 x 23.6 x 13.7 inches (W x H x D) | Weight | 45 kg (99 lbs) |
|
|