Instrument Database:
ANSYCO Analytische Systeme und Componenten GmbH - Voyager - Portable GC
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
ANSYCO Analytische Systeme und Componenten GmbH
|
Categories |
Chromatography: GC |
Portable gaschromatograph for the simultaneous determination of organic compounds at ppm and ppb-levels - Weight:6 kg
- Sample loop
- Pre-column (with back flushing)
- Three parallel analytical columns
- PID and/or ECD
- Concentration values given on display
- Isotherm column oven (55 - 80 °C)
- Identification and quantification via library function
- Storage of results
- Operation on mains and accumulator
- Sample pump
- Build-in gas reservoir
- 6 h independent operation time
- Air-, and Soil-air samples
- Method development with external PC
|
Specifications |
|
GCColumn temperature | up to 80 °C | Temperature stability | ~ 0,1 °C | Sampling pump | intern - 150 ml/min | Sample presentation | Sample loop or Injection | Detectors | PID, ECD | Sensitivity | 10 ppb benzene (PID), ca. 0,1 ppb CKW (ECD)) | Carrier gas | Nitrogen (Assay #1,4), synth.Air (Assay#2,5,6,7) | Peak integration | automated via internal processor | Concentration determination | automated via internal processor | Determination of retention times | automated via internal processor | Identification of compounds | automated via internal processor | Analysis-cycle time | 1 -90 minutes | Output of chromatogram | via RS232 auf | Data-storage capacity | 40 chromatograms | Dimensions | 39 x 27 x 15 (LxBxH) in cmT | Weioght | ca. 6,8 kg | Carrier gas | N2 or snth. Air capacity of reservoir sufficient for 8 h operation | Accumulator
| Ni/Cd (8 hours operation) | Power
| 220 VAC, 12 V DC 12 VDC Battery (for 6-8 hours operation) | Temperature | 0 - 40 °C | Ex-Certificate | EEx ibm IIC T4 |
|
|