Dionex Corporation - UltiMate™ 3000 Autosampler
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
Dionex Corporation
|
Categories |
Automation: Autosampler (LC) |
The Analytical Autosampler is an essential component of any Ultimate™ 3000 HPLC Analytical System and performs rapid, accurate, and precise injections without carryover between sample injections. The Analytical Autosampler uses an in-line, split-loop injection principle where the needle and sample loop are integral parts of the high pressure fluidic path to ensure 100% transfer of the sample volume. The Ultimate 3000 Analytical Autosamplers are ideally suited for routine analysis in pharmaceutical, food and beverage, and environmental laboratories, and are available in a standard and thermostatted version.
- Robust, dependable performance with low operating and maintenance costs
- Peak area precision is typically <0.15% RSD for 5-μL injections [more]
- Combination of a high-precision drive mechanism and the in-line, split-loop injection principle routinely delivers injector linearity values of r2=0.9990 from 5 – 80 μL [more]
- Volumetric accuracy better than ±0.5% for 20, 50, and 90 μL injections guarantees consistent results [more]
- Supports an injection volume range from 1 to 100 μL suitable for most applications, but can be extended to 2500 μL for added flexibility
- Well plate compatible sample trays paired with low injection cycle times (<15 s for 5 μL) support high throughput applications
|
Specifications |
|
sample capacity | Dependent on sample trays: 15 x 10 mL 66 x 4 mL 120 x 1.8 mL 3 x 96 normal and deep well plates 3 x 384 deep well plates | Injection methods | In-line split-loop | Injection volume | 1–90 µL (settable: 0.1–100 µL) | Injection volume accuracy | ±0.5% at 50 and 90 µL | Injection volume precision | <0.25% RSD at 5 µL (typically <0.15% RSD) | Linearity | Corr. coeff. >0.9999, RSD <0.5% at 5–80 µL | Needle wash | Active external needle wash | Carryover | <0.005% for caffeine with external wash | Injection cycle time | <15 s for 5 µL | Optional sample thermostatting | 4–45°C, at least 22 °C below ambient | Sample temperature accuracy | ±2 °C | Dimensions (h x w x d) | 40 x 42 x 51 cm (16 x 16.5 x 20 in.) | PC connection | USB; USB hub with 3 sockets integrated |
|
|