Instrument Database:
SPECTRO Analytical Instruments Inc. - ARCOS - Simultaneous ICP-OES Spectrometer
|
|
|
|
|
|
|
Year of introduction |
2007 |
Status |
available |
Company |
SPECTRO Analytical Instruments Inc.
|
Categories |
Spectrometer ( Atom. ): AES: ICP-AES |
Its appearance alone sends out a definite signal: The SPECTRO ARCOS is different from conventional ICP spectrometers. The prominent left side of the instrument houses the unique optical system – a patented major development with unbeatable resolution, accuracy and stability.
The SPECTRO ARCOS can be outfitted with an interface for either axial or radial plasma observation. The new CCD optic system with a Paschen-Runge mount assembly delivers a matchless resolution of 8.5 picometer in the wavelength range from 130 to 340 nm.
Additional SPECTRO ARCOS strengths include the new, extremely robust, free-running generator with a ceramic tube and solid-state power supply, the low maintenance and economical UV-Plus system for high VUV transparency down to 130 nm, as well as the fast readout system with a dynamic range of eight decades and a transient measuring frequency of up to 10 Hz. The patented ICAL system logic automatically monitors operation of the SPECTRO ARCOS guaranteeing continuous optimum operating conditions.
A high degree of flexibility, clear structure and simple operation characterize the “Smart Analyzer Vision” graphical user interface. Whether defining norm-conforming measurement procedures, quantitatively reprocessing stored spectra or creating measurement methods, operators appreciate the straightforward and easily accessible functions.
|
Specifications |
|
Dimensions (w x d x h in cm)
| 161.0 x 75.3x 107.4
| Weight (kg)
| 250
| Power
| 230 VAC ± 5%, 50/60 Hz – approx. 4.5 KVA power consumption – 30-32 A instrument required line protection (slow blow fuse)
| Spectrometer
| 0.75-m Paschen-Runge mount; Argon filled and recirculated vor UV transparency; | Grating | Holographic master grating made from Zerodur; 2 x 3600, 1 x 1800 lines/mm
| Optics
| thermally stabilized at 15 °C+- 0.5 °C | Wavelength range | 130-770 nm
| Wavelength drive | | Resolution | 8.5 pm (130-340 nm) 15 pm (340-770 nm)
| Detector | 29 linear CCD detectors; 3648 pixels per array
| Detection mode
| | Read out | Parallel read-out; shortest read-out time : 1ms |
| | ICP generator
| free-running oscillator; solid-state power supply; ceramic tube; air-cooled; Power efficiency: >70 %, power stability <0.1 % variation – Automatic plasma ignition – Stand-by mode (low power, low argon consumption) – Fully computer controlled | requency (MHz) | 27.11 | Power (Watt)
| 700-1700
| Load coil
| | Torch
| axially or radially viewed (optional)
| Argon flow (L/min)
| |
| | Sample introduction | four-channel peristaltic pump;
| Operating system and software
| SMART ANALYZER VISION
| Computer
| | Short term precision (RSD for 10 replicates)
| | Long term precision (RSD for 4 hours)
| | Sensitivity
| | Detection limits (3s, µg/L) | | Accessories
| | Special features
| |
|
|