Instrument Database:
LUMEX Ltd. - InfraLUM® FT-02 Fourier-Transform Spectrometer
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
LUMEX Ltd.
|
Categories |
Spectrometer ( Molec. ): FTIR |
InfraLUM® FT-02 is a multipurpose mid-IR FT spectrometer designed for laboratory use. The instrument is equipped with a data acquisition and processing system operating with a dedicated PC SpectraLUM® analytical software package. Description - Full-size sample compartment
- LUMEX proprietary DCE interferometer
- Permanently aligned optics
- Sealed and desiccated optics
- Shortened optical beam path
- User-friendly software
- Advanced data manipulation
- Standard accessories applicable
- Built-in diagnostics
Design features A proprietary "double cat's eye" design of the interferometer, which is vibration- and misalignment-proof, ensures high measurement stability, accuracy and precision. The Fourier transform of the detected interferogram reconstructs the IR spectrum throughout the spectral range. Fast interferogram detection (0.3 - 2.5 s, depending on resolution) allows multiple measurements to be made during a short period of time, which improves the signal-to-noise ratio. The large sample compartment can accommodate all conventional and non-standard large cells and accessories. The data acquisition and processing system, operating under Windows® 95/98 operation system, provides simple control and allows even unskilled personnel to easily master the instrument operation.
|
Specifications |
|
PERFORMANCE CHARACTERISTICS AND SPECIFICATIONS
Spectral range - total | 7500 - 350 cm-1
| Spectral range - with validated metrological characteristics | 4500 - 550 cm-1 | Resolution | 1, 2, 4, 8, and 16 cm-1
| Wavenumber accuracy | 0.05 cm-1 | Photometric accuracy | < 0.1% T
| Nonuniformity of a 100-% line | < 0.4% T
| Radiation source | ceramic | Signal-to-noise ratio* | 8000 | Detector | pyroelectric | Beamsplitter | Ge/KBr (Ge/ZnSe) | Power requirements | 110/220 V, 60 W
| Size and weight | 580x515x295 mm, 37 kg | | |
* For an RMS deviation of a 100-% line in the 2000-2200 cm-1 spectral region with a resolution of 4 cm-1, the measurement time is 1 min.
|
|