Instrument Database:
Teledyne CETAC Technologies - LSX-213 G2+ Laser Ablation System
|
|
|
|
|
|
|
Year of introduction |
2013 |
Status |
available |
Company |
Teledyne CETAC Technologies
|
Categories |
Sample introduction: Laser ablation |
The LSX-213 G2+ system features the latest in optical design technology to target areas of interest. Built on the rugged and field proven LSX platform, the Light Engine inside the LSX-213 is a hermetically sealed, MIL-SPEC laser head affording unmatched stability and user serviceability. Coupled with precision mounted optical components on a vibration isolated optical bench, the system can be used on multiple analytical instruments at multiple locations with confidence and ease. Simple, elegant component layout leads to improved stability and ease of maintenance. The laser ablation platform incorporates an open architecture sample area that enables flexibility for a wide range of specialized and custom sample cells. Spot sizes ranging from 4 μm to 200 μm are generated using aperture imaging technology, and are easily set using software control. From power, spot size and scan speed to control over timing and communication with your spectrometer, Teledyne CETAC Technologies is focused on making your job easier; simply Target, Ablate, and Analyze. The LSX-213 incorporates the latest in high resolution hardware and optical design technology, providing the user with excellent optical resolution and a wide field of view. Images are of startling quality, and high intensity lighting arrays above and below the sample produce extremely clear images even at high zoom.
|
Specifications |
|
Laser Cabinet | Frequency quintupled 213 nm Q-switched Nd:YAG laser with > 4 mJ source pulse energy and < 5 ns pulse width | | < 3% pulse to pulse stability throughout | | Flat top energy profile | | Mechanically and thermally isolated laser beam delivery and viewing path for maximum stability | | Variable output energy and variable pulse repetition frequency (1 – 20 Hz) | | Variable laser spot size (4 – 200 μm) | | Fully sealed laser head with user changeable flashlamp | Viewing Optics and Video | Computer controlled, continuously adjustable parfocal video microscope (2.6x – 32.5x optical zoom range) | | Wide field of view (> 6 mm) at high optical resolution (< 2 µm) | | High intensity LED lighting arrays | | Motorized rotating polarizer | Sampling System | Open Architecture for maximum flexibility | | Optional HelEx Active 2-Volume Cell | | High precision XY stages (0.16 µm/step) |
|
|