Technical Data
Maximum operating pressure: 6000 psi (410 atmospheres)
Flow rate range: 0.01 to 10.0 mL/min in 0.01 mL/min steps between 0.00 and 1.00 mL/min, and 0.1 mL/min steps between 1.0 and 10.0 mL/min
Flow precision: 0.3% RSD at 1.0 mL/min MeOH/water
Flow accuracy: ± 0.5% at 1.0 mL/min isopropyl alcohol
Compositional accuracy: ± 0.5% absolute
Compositional precision: ± 0.1% absolute
Delay volume: 700 µL + 1 µL/ATM pressure
Pressure pulsation: Less than 2% at 150 atmospheres (1 mL/min with 100% isopropyl alchohol)
Proportioning modes: 1, 2, or 4-stroke (user selectable)
Mixing: High pressure, static or dynamic mixing modes
Display: 40 Characters, double-line, back-lit LCD
Interface
Contact closure outputs: Six time-programmable external event relays, and three synchronization signals: READY, START, and FAULT
Contact closure inputs: Three synchronization signals: READY, START, and FAULT
Bus communication: The 220/230/240 solvent delivery systems communicate with the Star Workstation by means of the IEEE 488 Bus (GPIB)
Environment
Operates within specifications from 10 °C to 35 °C. Non-operational storage from
-20 °C to 65 °C. Operates in relative humidity from 5% to 95%.
Dimensions: 29.2 cm (w) x 29.2 cm (h) x 50.2 cm (d) 11.5 in. (w) x 11.5 in. (h) x 19.75 in. (d)
Weight: 27.7 Kg (61 lbs.)
Cat. No. Description
0393613005 ProStar 220 Isocratic Solvent Delivery Module
0393613001 ProStar 230 Ternary Gradient Solvent Delivery Module
0393613002 ProStar 230 Inert Ternary Gradient Solvent Delivery Module
0393613003 ProStar 240 Quaternary Solvent Gradient Delivery Module
0393613004 ProStar 240 Inert Quaternary Solvent Gradient Delivery Module