Instrument Database:
Hewlett-Packard - HP 1050 Quaternary HPLC Pump ( Agilent Technologies Inc. )
| |
|

|
| |
|
| |
|
| Year of introduction |
|
| Status |
historical ( out of sale ) |
| Company |
Hewlett-Packard
|
| Categories |
Chromatography: LC/HPLC |
|
| Specifications |
|
|
Specifications: | Flow range | Setpoints from 0.001 to 9.999 ml/min in 0.001 ml/min increments | | Flow precision | < 0.3% RSD (typically < 0.15 %), based on retention time at 0.5 ml/min and 2.5 ml/min | | Pressure | Operating range from 0-400 bar up to 5 ml/min; from 0-200 bar up to 10 ml/min | | Compressibility compensation | User-selectable, based on mobile phase compressibility | | Recommended pH range | 2.3 to 12.5 | | Gradient formation | Low pressure quaternary mixing/gradient capability using proprietary high-speed proportioning valve | | Composition Range | 0 to 100% in 0.1% increments from four independent channels | | Methods | Battery-backed storage of up to 10 methods | | Analog Output | For pressure monitoring, 2 mV/bar |
|
|