Instrument Database:
Hewlett-Packard - HP 1100 HPLC System with Binary Pump, Autosampler and Fluorescence Detector ( Agilent Technologies Inc. )
|
|
|
|
|
|
|
Year of introduction |
|
Status |
historical ( out of sale ) |
Company |
Hewlett-Packard
|
Categories |
Chromatography: LC/HPLC |
|
Specifications |
|
1100 Series Fluorescence Detector Specifications: Detector Type | Multi-signal Fluorescence detector with rapid on-line scanning capabilities and spectral data analysis | Light source | Xenon Flash Lamp, normal mode 20W, economy mode 5W | Excitation Monochromator | Range: 200 nm - 700 nm and zero-order Bandwidth: 20 nm (fixed) Monochromator: concave holographic grating, F/1.6, blaze: 300nm | Emission Monochromator | Range: 280 nm - 900 nm and zero order Bandwidth: 20 nm (fixed) Monochromator: concave holographic grating, F/1.6, blaze: 400nm | Pulse Frequency | 296 Hz for single signal mode 74 Hz for spectral mode | Reference System | in-line excitation measurement | Wavelength characteristics | Repeatability: +/- 0.2 nm Accuracy: +/- 3 nm setting | Flow cells | Standard: 8 µl volume and 20 bar (2 MPa) pressure maximum, quartz | Analog outputs | Recorder/Integrator: 100 mV or 1 V, two outputs |
|
|