Instrument Database:
asi - Analytical Scientific Instruments - PrimeLine™ Isocratic Pump Model 500
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
asi - Analytical Scientific Instruments
|
Categories |
Chromatography: LC-Pump Chromatography: LC-Pump: Isocratic |
The ASI HPLC pump is designed with the chromatographer in mind. It can be used as a reliable low cost isocratic pump or integrated into a gradient HPLC System. In either case, the pump incorporates several key features which make it truly one of a kind in the world of HPLC Pumps.
Self-Priming
The pump will prime itself with any solvent and at any flow rate. Turn on the pump and open the purge valve, the pump will prime itself even from a completely dry start. You will never need to use syringes to prime it.
Patented Floating Pump Seal
The seal actually floats so that it is always perfectly aligned with the piston. Our tests have proved that the average life with this new patented design is more than 5,000 hours of continuous operation at 6,000 PSI and 10 mL/min. with pure water.
Intuitive Software Control
The control is designed with dedicated keys that allow even a novice to turn on the pump and create methods. An easy to read 4-line x 20-character LCD display allows the user to monitor several pump parameters from the same screen. The model 501 can be controlled from its own keypad or from a Model 500 gradient controller. When configured as a gradient HPLC system users can program and store up to 60 methods with 20 lines each. Methods may be linked or run independently.
Solvent Compressibility Compensation
The user may select from several different solvents from the on board library. The control automatically compensates for compressibility of the selected solvent.
Easy Maintenance
All user serviceable components are easily accessed from the front panel. Once the head assembly is removed, the pistons and seals can be replaced quickly and easily without the use of tools.
Pressure Sensor
Bonded strain gage with flow through design and digital output
Prime/Pure Assembly Allows for quick and easy solvent change
|
Specifications |
|
The Model 500 has been designed to convert the Model
501 from an isocratic pump to a gradient HPLC system. Full operation
and gradient control programming is achieved through the keypad on the
front panel of the Model 500. The Model 500 includes a high pressure
integral static mixer, connecting tubing, and all the accessories necessary
to quickly convert the Model 501 to gradient operation.
Model 500/501 Specifications: Analytical Pump Head Specifications Flow Rate: Settable from .001 mL/Minute to 10 mL/Minute Maximum Pressure Setting: 6,000 psi (Default = 5,400 psi) Maximum Pressure Cap.: 6,000 psi at 10 mL/minute as indicated on the pump display panel. Accuracy: ± 1 % or ± 2 microliter/min. whichever is greater Precision: 0.25 % from 0.1 mL/min. to 10 mL/min. at 20 degrees C Dimensions: 10-¼” W x 14” D x 6-½” H Weight: 15 lbs. (6.8 kg) Power input: 85 to 264 VAC, 47 to 63 Hz, 60 watts Model 500/501 Specifications: Semi-Prep Pump Head Flow Rate: Settable to .010 mL/Minute to 20 mL/Minute Maximum Pressure Setting: 3,000 psi (Default = 2,500 psi) Maximum Pressure Cap.: 3,000 psi at 20 mL/minute as indicated on the pump display panel. Accuracy: ± 1% or ± 4 microliter/min., whichever is greater Precision: 0.25% from 1.0 mL/min. to 20 mL/min. at 20 degrees C Dimensions: 10-¼” W x 14” D x 6-½” H Weight: 15 lbs. (6.8 kg) Power input: 85 to 264 VAC, 47 to 63 Hz, 60 watts Model 500/501 Specifications: Prep Pump Head Flow Rate: Settable to 100 mL/Minute to 80 mL/Minute Maximum Pressure Setting: 1,440 psi (Default = 1,200 psi) Maximum Pressure Cap.: 1,000 psi at 80 mL/minute as indicated on the pump display panel. Accuracy: ± 2% or ± 20 microliter/min., whichever is greater Precision: 0.5% from 10.0 mL/min. to 80 mL/min. at 20 degrees C Dimensions: 10-¼” W x 14” D x 6-½” H Weight: 15 lbs. (6.8 kg) Power input: 85 to 264 VAC, 47 to 63 Hz, 60 watts
|
|