Instrument Database:
Dionex Corporation - EDA50A Electrochemical Detector
|
|
No picture !!
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
Dionex Corporation
|
Categories |
Chromatography: LC-Detector: Electrochemical |
For excellent sensitivity and selectivity for a wide range of compounds, use the versatile ED50A Electrochemical Detector. Get the benefits of performance and versatility provided through the following unique features:
- Advanced amperometric cell design that incorporates a 1-mm-diameter working electrode for a high signal-to-noise ratio
- Four working electrode options for amperometric cell: gold, platinum, silver, or glassy carbon
- Low-volume amperometric cell that minimizes dispersion
- Corrosion-resistant titanium flow path and counter electrode
- Combination pH-Ag/AgCl reference electrode that minimizes baseline shift with pH gradients
- Optional sodium reference electrode
- Easy remote mounting of either conductivity or amperometry cell
- Conductivity cell geometry that provides excellent accuracy at both low and high conductivities
- Low electrode mass in the conductivity cell that allows for effective temperature compensation for improved reproducibility
- Optional DS3 detection stabilizer provides conductivity measurements nearly impervious to temperature variation for improved reproducibility
- Built-in control for the SRS® Self-Regenerating Suppressor, the most advanced suppressor for eliminating eluent detector background and providing high signal-to-noise
|
Specifications |
|
CONDUCTIVITY MODE: Noise ±0.1 nS when background conductivity is 0–10 µS ±1 nS when background conductivity is 10–100 µS ±10 nS when background conductivity is 100–1000 µS
Cell drive 8 kHz square wave
Linearity 1% at 1 mS
Resolution 0.1 nS
Analog output 11 separate ranges from 0.01 to 3000 µS
Flow cell Electrodes: Passivated 316 stainless steel microelectrodes Cell Body: Chemically inert polymeric material Cell Volume: 1.0 µL Maximum Operating Pressure: 2 MPa (300 psi)
Temperature compensation Selectable compensation between 0.0 and 3.0% per °C
AMPEROMETRY MODE: Noise dc amperometry: 1 pA Integrated amperometry: 20 pC
Potential settings –2.00 to 2.00 V in 0.01-V increments
Analog output dc amperometry: 50 pA–300 µA Integrated amperometry: 50 pC–200 µC
Flow cell Working Electrode (interchangeable): gold, platinum, silver, or glassy carbon Counter Electrode: titanium Reference Electrode: pH-Ag/AgCl combination (standard); Na (optional) Wetted Materials: PEEK, titanium, Kel-F®, polyetherimide, EPR, glass, high-density polyethylene (HDPE) Cell Volume at Working Electrode: <0.5 µL Maximum Operating Pressure: 0.7 MPa (100 psi)
GENERAL SPECS: Electronics Microprocessor-controlled digital signal processing
Autoranging Autoranging digital conductivity signal monitoring with Chromeleon software through DX-LAN
Analog output User-selectable full-scale recorder output of 10, 100, or 1000 mV
Control modes Local or remote using relay closures, TTL, or by Chromeleon® software through DX-LAN
|
|