Instrument Database:
EquiLab Inc. - NE 010 Interlock
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
EquiLab Inc.
|
Categories |
Separation: Centrifuges |
This range consists of two models constructed with a steel body with fourplace swing out rotor complete with removable aluminum buckets. Each version has safety interlock, which prevents the lid being opened until the rotor is at rest.
An additional safety feature ensures that in the case of power failure the lid remains in the locked position. The product can only be operated if the lid is correctly closed and locked in position.
Features include:
- Control panel has a series of indicator lamps
- for rotor on, lid locked and rotor run down
- Maintaining finger pressure on the push button operates the centrifuge and locks the lid in position
- Releasing finger pressure de-energies the rotor, but the lid remains locked until the rotor is at rest
- Timer (NE010GTH) allows variable rotor operation up to 30 min.
|
Specifications |
|
Operation Rotor activated push button (NE010G/I) Timer:0-30 min. (NE010GT/I)
Max. speed 2800 rpm
Rotor type Swing out or angle head
Tube capacity 15 ml
Max. number of tubes 4
Tube dimensions Ř 16 x 110 mm Ř10 x 75 mm with adaptors
Max. radius 135 mm
Max. RCF 1200 g
Overall dimensions Lid closed: 361w x 400d x 320h mm Lid open: 361w x 415d x 720h mm
Rating 100 W
Voltage 230V, 50 Hz
|
|