Instrument Database:
Mettler Toledo AG - Berger SFC™ Analytix/MS
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
Mettler Toledo AG
|
Categories |
Chromatography: SFC |
The Berger SFC™ Analytix/MS extends the offering of the Berger SFC™ Analytix system by integrating the Waters® ZQ™ Mass Detector, thereby allowing the user to harvest more data. The BERGER SFCTM Analytix/MS system components include: - SFC Dual Pump Module
- Leap HTC PAL Autosampler
- Column Oven with 2 or 6 Column-Switching Valve
- DAD or VWD
- Waters® ZQTM Mass Detector with APCI and ESI Sources
- MS Make-up Pump
- SFC MasswareTM Software
|
Specifications |
|
System Specifications | Flow Rate | 0 to 10 ml per minute | Detectors | VWD, DAD, Waters ZQ MS | Columns | 4.6 mm recommended | Pressure | 0-400 bar | Fluid Control Specifications | External Dimensions | 32 cm wide, 41 cm high, 53 cm deep, 43 kg weight | Voltage | 100/120/220/240 | Frequency | 50 or 60 Hz (±5%) | Compressible Fluid Pump | Hydraulic System | Reciprocating dual pistons in series, variable stroke, variable frequency | Compressed fluid flow range | Setpoints from 0.001 to 9.99 ml/min | Flow Ramps | 3 | Liquid Modifier Pump | Hydraulic system | Reciprocating dual pistons in series, variable stroke, variable frequency | Modifier Concentration | 0 to 100% in 0.1% increments | Total System Flow | 10ml/min | Pressure Control | Pressure range | Up to 423 bar (0 to 5880 psi) | Typical pressure ripple | <0.5 bar peak-to-peak measured at the outlet of a packed column | Pressure ramp | 10 at up to 99 bar/min | Pressure transducer | Linearity better than ±0.25% calibrated to <± 0.5 bar |
|
|