Instrument Database:
VG Isotopes - PlasmaQuad (PQ I) ( Thermo Scientific )
|
|
|
|
|
|
|
Year of introduction |
1984 |
Status |
historical ( out of service ) |
Company |
VG Isotopes
|
Categories |
Mass spectrometer: ICP-MS |
First generation quadrupole ICP-MS instrument. The PQ1 was developed in 1984 by VG Elemental and was the first commercially available ICPMS. Standing approximately 6ft high and 9ft wide it has quite a room filling presence. The plasma RF was generated from a 2KW Henry generator and the vacuum system was made up of Blazer diffusion pumps and Edward's rotary pumps. The torchbox moved on a pneumatic piston driven carriage and it had the 12-12 quadrupole system. Sensitivity was in the range of 1 to 4 mcps/ppm and the instrument was controlled by DOS software.
|
Specifications |
|
Dimensions (w x d x h in cm) | 216 x 65 x 160 (spectrometer) 60 x 65 x 112 (RF generator)
| Weight (kg)
| 550
| Power
| 3 phase, 20 A each, 415 V, 10 kVA | Mass filter
| quadrupole
| Mass range
| 0-300 amu | Resolution | 1M and 3M over the mass range | Detector
|
| Dynamic range (orders of magnitude)
| 6 | Vacuum system | 3 stage differentially pumped system; diffusion pumps on expansion and analyser stage | ICP generator
| Frequency-stabilized | Frequency (MHz) | 27.12 | Power (Watt)
| 1500 (2000/2500 optional) | Load coil |
| Torch
|
| Argon flow (L/min)
| 13 | Sample introduction
|
| Operating system and software
| DOS
| Computer
|
| Short term precision (RSD for 10 replicates)
|
| Long term precision (RSD for 4 hours)
|
| Sensitivity (Mcps/ppm) | 1 to 4 | Detection limits (3s, ng/L)
|
| Isotope ratio precision (RSD for Ag-107/Ag-109)
|
| Doubly charged Ba ions
| < 2% | Ce Oxide ions | < 0.5% | Accessories | laser ablation; 200 position autosampler | Special features |
|
|
Related links:
|