Instrument Database:
Knauer GmbH - Smartline PDA Detector 2800 UV
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
Knauer GmbH
|
Categories |
Chromatography: LC-Detector: Diode Array |
With a deuterium lamp for highly sensitive measurements between 190 and 600 nm The Smartline Diode Array Detector 2800 with a UV light source supports a wavelength range of 190 to 600 nm. Aside of having an enlarged linearity range, this detector is distinguished by having an outstanding signal to noise ratio. Work with spectra libraries is supported by the detector which makes the inspection of peak purity possible. Maintenance and exchange of the measuring cell is made simple by having direct access to it from the front panel.
|
Specifications |
|
Diode Array 512 diodes in a ceramic chassis
Wavelength Range 190-600 nm
Light Source Deuterium lamp
Spectral Resolution 2.4 nm (0.8 nm pixel distance)
Wavelength Accuracy < 0.5 nm
Wavelength Reproducibility < 0.1 nm
Signal-to-Noise Ratio 1 x 10-5 AU, according to ASTM, optical wavelength 10 mm, 1 ml/min methanol, λ= 254 nm, Δλ= 4 nm (±2 nm)
Linearity 0-2.0 AU
Drift < 5 x 10-4 AU/h, according to ASTM E1657-94
Technical Parameters Digital Control and Output: Ethernet, Analog output: 2 x ± 10 V, Power Supply: 115/230 V, 50-60 Hz, 100 W
Dimensions / Weight 226 x 185 x 390 mm (W x H x D) / 8.5 kg
|
|