Instrument Database:
getspec.com - getSpec NIR1.7-256-PC
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
getspec.com
|
Categories |
Spectrometer ( Molec. ): NIR |
The getSpec NIR1.7-256-PC with USB interface is a complete NIR spectrometer with a Peltier-cooled InGaAs diode array detector on a PC card. The spectrometer has no flexible or adjustable parts and is completely maintenance-free. This guarantees a high degree of stability of the NIR calibrations models. High detector sensitivity and Peltier cooling of the InGaAs diode array also permit high speed measurements in the millisecond range.
Owing to the flexible fiberoptical coupling, the spectrometer can be used for various different applications in process analysis, mobile analysis or quality assurance. In connection with chemometric methods, it is well suited for quantitative as well as qualitative analysis.
In addition to the SPEC32 software that is included in delivery, a GRAMS®/32 package and a MyInstrument driver are available optionally. Using the driver the spectrometer can directly run under GRAMS®/32 software. This allows you to build your chemometric models for quantitative or qualitative analysis easily.
Should you have questions concerning the technical specification; our application scientists will be pleased to consult you.
|
Specifications |
|
Spectral range 900 nm - 1700 nm Customized ranges available
Detector Single stage TE-cooled InGaAs diode array (25°C from Tamb) < 2% dropout pixel, 0% dropout on inquiry
Pixel size ( width x height) 50 µm x 500 µm 25 µm x 500 µm
Pixel number 256 512
Spectral dispersion Ca. 3.12 nm / pixel Ca. 1.56 nm / pixel
Spectral resolution Ca. 6 nm Ca. 3 nm
Stray light < 0.2%
Wavelength stability < 0.05 nm over 48 hours
Fiber optic connector SMA connector, internal 400µm-LOH optical fiber
Integration time 10 µs - 8 s
Detector sensitivity 500 A/D counts/sec/pW @ 1200 nm 500 A/D counts/sec/pW @ 1450 nm 150 A/D counts/sec/pW @ 1650 nm
Photometric repeatability < 50 µAU, Variation between 100 data points, Each point consists of 1000 samples (8 s accumulation)
Photometric stability < 0.5 mAU over 48 hours
Noise level 13 µAU RMS (approx. 2 counts for 1024 point sample, 8 s accumulation, reference of 65,000 counts)
Signal-to-noise ratio 32000:1
A/D converter 16 bit, 330 kHz
Electrical interfaces Trigger input TTL, 2 digital TTL outputs
PC interface USB
Power requirements 230 V / 50 Hz (12 VDC available on request)
Size 300 x 90 x 190 mm
Weight 3.35 kg
|
|