Instrument Database:
Beijing Rayleigh Analytical Instrument Corporation (BRAIC) - WLD-4C Optical Emission Spectrometer
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
Beijing Rayleigh Analytical Instrument Corporation (BRAIC)
|
Categories |
Spectrometer ( Atom. ): AES: Arc/Spark |
Features:
- Reasonable design and perfect performance;
- Rapid analysis, perfect repeatability and stability;
- High integrated data collection and control system, high automatic degree;
- Used for many kinds of matrix analysis: Fe, Co, Ni, Ti, Cu, Al, Pb, Mg, Zn, Sn etc.
- High repeatability & stability excitation source is adopted. Excitation frequency can
- change in 150 ~ 600 Hz range. Select it based on analysis material user analyzes, to reach best analysis effect.
- Computer software is user-friendly and easy to be operated;
- Database is built in. User can transfer data through the internet, convenient and quick.
- Shake-protection performance is strong, and there is no need to make shake-protection base.
- Local constant-temperature technique is adopted to ensure the normal operation,minimizing
- requirement to environmental conditions.
- Operations and maintenance are easy and convenient, and maintenance cost is low.
|
Specifications |
|
Spark Source Power supply: AC 220v ±1% 50 Hz Input power: 1.0 kVA Charge capacitance: Pre-burn 7 µf, Expose 1.5 µf Discharge Inductance: Pre-burn 35 µh; Expose 70μh Peak current: Pre-burn 100A; Expose 35A Peak voltage: 300V Ignite circuit: Amplitude of pulse +15 KV Auxiliary gap 5mm Discharging frequency: 300 Hz/150Hz or 400Hz/200Hz
Optical Unit Spectrum range: 175∼450 nm Optical channels: up to 36 channels Concaved grating: Focus length 750 nm Groove density 2400 lines /mm Ruled area 30 mm × 50 mm Blaze wavelength 300 nm Linear dispersion 0.55nm/mm Width of entrance slit: 20 µm Width of exit slit: 50 µm, 75 μm Constant temperature: 30±0.1 ℃ Gas supply pressure: 0.01~0.02MPa
Readout Unit Readout mode: Subsection integration Repeatability RSD ≤ 0.2% Integrate capacitance 0.047 µf PMT HV power supplies Voltage -1000v Stability ≤0.5% (8 hours) HV adjustment Change through computer software
Data Processing Unit Hardware: Computer embedded system, CPU main frequency 300MHz Display industrial 12.1 inch LCD Printer 24 pins Software (Please refer to WLD-4C Multi-channel Optical Emission Spectrometer Software Guide Book.)
Physical Data Mainframe: 1500 (l) ×1100 (w) × 1200 (h) mm3 Weight: about 450 kg
|
|