| Sample Injection Unit |
| Sample Injection System | Total sample injection, variable sample injection system (zero sample loss during injection) |
| Settable Range Of Injection Volumes | 0.1 to 100µL (standard), or 1 to 2000µL(option) (0.1µL steps from 0.1 to 0.9µL, and 1µL steps from 1 to 2000µL) |
| Sample Processing Capability | With 1-mL vials: 350 With 1.50mL vials: 210(SIL-HTa), 140(SIL-HTc) With 4-mL vials:100 With micro plates:4 (96 wells: 384, 384 wells: 1536) With deep-well plates: 4 (96 wells: 384, 384 wells: 1536) |
| Injection Volume Repeatability | RSD: 0.3% max. (at 10µL injection) |
| Cross Contamination | 0.01% max. (for analysis of naphthalene) |
| Injection Volume Accuracy | Less than +/-1% (50µL) |
| Injection Volume Accuracy Correction | Possible |
| Number Of Injections | 1-99 times per sample |
| Settting Of Analysis time | 0.01-9999.9 (0.01 minute increments to 999.099min.,0.1 minute increments over 1000min.) |
| Number Of Steps For Injection Conditions | Max 102 steps |
| Sample Suction Speed | 0.1 - 15µL/sec (0.1µL/s step for 0.1-0.9µL/sec., 1µL/s step for over 1µL/sec.) |
| Rinsing Liquid Suction Speed | 1 - 35µL/sec (1µL/s step) |
| Maximum Pressure | Max. 20.0Mpa (Standard pipings of sample flowpath is PEEK) |
| Sample Cooler (SIL-HTc only) |
| Method | Direct cooling method (at a setting of 4°C, room temperature must be less than 30°C and humidity less than 70%) |
| Temperature setting range | 4 to 40° (Cooling down to 4°C for room temperature up to 30°C and humidity up to 70%) |
| Operating Ambient Temperature Range | 4 to 35°C |
| System Controller |
| Display | LCD with back light (320 x 240 pixels) |
| Memory Media | 3.5 inch floppy disk drive, disk (2HD) |
| Input/Output Terminals | External start input: 1, Error input: 3, External output(EVENT 1,2,3,4): 4, Output for external control (AC Remote):1, Optical link: (for Chromatopac), 8 (for LC unit control), RS-232C: 1 (for PC connection) |
(configuration of condition setting files) No. of analysis file No. of fraction collector files Analysis sequence files | 20 files (parameter, time program) 10 files (parameter, time program) 102 steps |
| Pump Control | Solvent delivery models: isocratic, high-pressure gradient, low-pressure gradient, constant pressure solvent delivery Programs: flow volume, pressure, concentration, upper pressure limit, lower pressure limit, linear step, index function (all are multiple steps) |
| Autosampler Control | Sample injection volume, number of repeated analyses, analysis time, analysis file unmber, etc |
| Column Oven Control | Oven temperature, upper temperature limit |
| Detector Control | Detection wavelength, range, time constant, lamp ON/OFF, scanning conditions, etc. |
| Self-diagnostic and safety measures | Monitoring abnormalities through checks of memory, upper and lower pressure limits oven temperature upper limit, lamp current and motor speeds |
|
| Operating pH Range | pH: 1 to 14 standard |
| Wetted Surface Material | SUS316, PEEK, Ceramics |
| Operating Temperature Range | 4 to 35°C |
| Operating Humidity Range | up to 70% |
| Power Supply | AC100V, 150VA (SIL-HTa), 300VA (SIL-HTc), 50/60Hz |
| Size | 540 (W) x 500 (D) x 415 (H) mm |
| Weight | 40kg (SIL-HTa), 45kg (SIL-HTc) |