Instrument Database:
HORIBA Jobin Yvon - ACTIVA
|
|
|
|
|
|
|
Year of introduction |
2002 |
Status |
historical ( out of sale ) |
Company |
HORIBA Jobin Yvon
|
Categories |
Spectrometer ( Atom. ): AES: ICP-AES |
Maximize your productivity and performance. ACTIVA™ builds on A Platform for the Future™ maintaining the strengths of Jobin Yvon ICP spectrometers while incorporating advanced CCD technology to improve speed and analytical throughput offering: - Simultaneous analysis and background at multiple wavelengths within a *Wavelength Analytical View (WAV).
- Superb high resolution which remains constant within a WAV and across the spectrum from 160 to 800 nm.
- High performance mega pixel CCD detector tailored for spectroscopy.
Flexibility through the user selection of any wavelength from 160 to 800 nm. - High quality optics: large JY holographic gratings, high luminosity, negligible stray light and optical aberrations.
- Multi-WAV acquisition provides simultaneous analysis of 75 elements in less than 30 seconds.
|
Specifications |
|
Dimensions (w x d x h in cm)
| | Weight (kg)
| | Power
| Single phase, 3 kW, 30 A, 220/240 V, 50/60 Hz
| Spectrometer
| 640 mm Czerny-Turner monochromator, nitrogen purged
| Grating | two ion-etched holographic gratings 80 x 110 mm, used back-to-.back
I: 3600 lines/mm (160-510 nm) II: 2400 lines/mm (160-800 nm)
| Optics
| achromatic entrance imager
| Wavelength range | 165 - 800 nm
| Wavelength drive | Full range in 10 s; reproducibility +- 1 pm
| Resolution | I: 13 pm (first order)
II: 10 pm (second order)
| Detector | 2048 x 512 pixel back-illuminated CCD
| Detection mode
| | Read out | |
| | ICP generator
| frequency stabilized; reflected power control | Frequency (MHz) | 40.68
| Power (Watt)
| 2600
| Load coil
| | Torch
| exchangeable casette system for aqueous/organic/suspension/HF solutions radially viewed
| Argon flow (L/min)
| 12 (7 low-flow option)
|
| | Sample introduction | | Operating system and software
| Windows XP/NT compatible software | Computer
| | Short term precision (RSD for 10 replicates)
| | Long term precision (RSD for 4 hours)
| | Sensitivity
| | Detection limits (3s, µg/L) | | Accessories
| hydride generator; HF-resistant nebulizer and injector, ultrasonic nebulizer
| Special features
| |
|
|