Instrument Database:
Baird - ICP 2000 - Simultaneous Plasma Emission Spectrometer ( Thermo Scientific )
|
|
|
|
|
|
|
Year of introduction |
1990 |
Status |
historical ( out of service ) |
Company |
Baird
|
Categories |
Spectrometer ( Atom. ): AES: ICP-AES |
This is Baird's top-of-the-line plasma Instrument, designed for the determination of up to 60 elements by simultaneous ICP emission spectrometry. Engineered into this rugged selfcontained System are highresolution polychromator optics; a 40 MHz, solid state, RF generator with Iow-flow plasma torch; and easily-learned, menu-driven software.
|
Specifications |
|
Dimensions (w x d x h in cm)
| | Weight (kg)
| | Power
| | Spectrometer
| 1-m polychromator, Paschen-Runge mount; Shock-mounted, stress-relieved, vacuum-welded enclosure sealed to prevent contaminatjon. Matched coefficients of thermal expansion for dimensional stability despite temperature changes.
0.5-m monochromator (optional) ; Czerny-Turner mount; argon purged for wavelength coverage Irom 162 to 800 nm
| Grating | 1440 grooves/mm; 50 x 30 mm; blazed at 400 nm
| Optics
| Focusing lens, quartz window and front-surfaced, UV-coated mirror purged with argon to prevent contamination; up to 230 pre-aligned exit slits | Wavelength range | 345 to 780 nm in first order, 170 to 385 nm in second order
| Wavelength drive
| sine bar positioning for scanning monochromator; Computer-controlled translation device provides scanning over 0.84 nm in polychromator
| Resolution | 0.664 nm/mm in first order, 0.332 nm/mm in second order
| Detector | up to 60 PMTs, gain voltage under computer control
| Detection mode
| | Read out | |
| | ICP generator
| crystal controlled for 0.05% frequency stability, automatic impedance matching network
| Frequency (MHz) | 40.68
| Power (Watt)
| 1500
| Load coil
| | Torch
| One-pieee quartz torch, high/low flow. Optional two-piece HF-resistant torch. Precision X-Y carriage movement for varying observation zone from external controls
| Argon flow (L/min)
| |
| | Sample introduction | Concentric pneumatic glass nebulizer. Optional HF-resistant, high-solids, V-groove nebulizer. Glass spray chamber. Optional HF-resistant system
| Operating system and software
| Baird Plasmacomp V Software
| Computer
| IBM Personal System 2 Computer. High-resolution 12" color monitor. Dot matrix printer with 10" carriage
| Short term precision (RSD for 10 replicates)
| | Long term precision (RSD for 4 hours)
| | Sensitivity
| | Detection limits (3s, µg/L) | | Accessories
| 102 position autosampler; UDX nebulizer; automated hydride generator
| Special features
| |
|
|