Spectrometer ( Molec. ): UV/VIS |
Specifications |
|
- A wide wavelength range: 190 nm to 1100 nm.
- Scan speeds from 7.5 nm/min to 2880 nm/min
- Double beam optics for excellent long term stability, reference compensation and baseline correction.
- Basic methods for time drive, scan, wavelength program and concentration. Store up to 20 ready-to-run methods.
- Storage for preprogrammed methods, allowing transfer from instrument-to-instrument, lab-to-lab
- Simple keyboard layout for ease of use. HELP prompt.
- Security of operation, allowing some instrument functions to be disabled to prevent misuse.
- Connection to PC operating with PECSS software.
- Stray Light (2 sec resp time): < 0.03 T at 220 nm
- Wavelenght Accuracy: 0.3 nm at D2 peak 656.1 nm
- Wavelenght Reproducibility: O.1 nm
- Slit Width: 2 nm
- Photometric Accuracy: 0.005 A at 1 A
- Photometric Reproducibility: 0.002 A at 1A
- Photometric Range: -3 A to 3 A.* Photometric Stability: < 0.0003 A/h (500 nm, 1 sec resp)
|
|