Instrument Database:
Dionex Corporation - CD25A Conductivity Detector
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
Dionex Corporation
|
Categories |
Chromatography: LC-Detector: Conductometry |
Get the benefits of performance and versatility provided by the CD25 Conductivity Detector’s unique features:
- CD25 cell geometry provides efficient sweepout, low volume, low dispersion, and excellent accuracy at both low and high conductivities.
- Low electrode mass in the conductivity cell allows for effective temperature compensation for improved reproducibility.
- The conductivity cell is mounted remotely at the end of the suppressor or column, minimizing dead volume.
- The optional DS3 detection stabilizer provides conductivity measurements that are nearly impervious to temperature variation for improved reproducibility.
- Use built-in control for the SRS® Self-Regenerating Suppressor, the most advanced suppressor for eliminating eluent detector background and providing high signal-to-noise ratio.
|
Specifications |
|
Electronics type Microprocessor-controlled digital signal processing
Cell drive 8-kHz square wave
Linearity 1% at 1 mS
Resolution 0.1 nS
Full-scale analog output ranges Eleven separate ranges from 0.01 to 3000 µS
Electronic noise ±0.1 nS when background conductivity is 0–10 µS ±1 nS when background conductivity is 10–100 µS ±10 nS when background conductivity is 100–1000 µS
Filter Rise times from 0 to 10 s
Temperature compensation Selectable compensation between 0.0% and 3.0% per 1 °C
Analog signal output voltage User-selectable full-scale recorder output of 10, 100, or 1000 mV
Control modes Local or remote using relay closures, TTL, or by Chromeleon or PeakNet software through DX-LANTM
Autoranging Autoranging digital conductivity signal monitoring with Chromeleon or PeakNet software through DX-LAN
Conductivity flow cell Electrodes: Passivated 316 stainless steel microelectrodes Cell Body: Chemically inert polymeric material Cell Volume: 1.0 µL Maximum Operating Pressure: 2 MPa (300 psi)
|
|