Instrument Database:
Applied Research Laboratories (ARL) - Model 3560 - Simultaneous ICP-OES ( Thermo Scientific )
|
|
|
|
|
|
|
Year of introduction |
1982 |
Status |
historical ( out of service ) |
Company |
Applied Research Laboratories (ARL)
|
Categories |
Spectrometer ( Atom. ): AES: ICP-AES |
Offering simultaneous analysis for high sample throughput in routine process control operations, the 3560 OES is a polychromator-based system utilising a one meter Paschen-Runge spectrometer. Accurate, high speed determination of elemental concentration for up to 60 elements are the hallmarks of the ARl 3560. Utilising the latest microprocessor technology, a complete multi-elemental analysis can be accomplished in less than a minute -the ideal selection for process control and quality assurance.
|
Specifications |
|
Dimensions (w x d x h in cm)
| 208 x 102 x 160
| Weight (kg)
| 1000 | Power
| 230 Vac, 50-60 Hz, 4KVA maximum for spectrometer and RF generator. Line regulator required if fluctuations exceed + 5%.
| Spectrometer
| 1-m Paschen-Runge mount, vacuum and air version; thermostatization +- 0.1 °C allows 18-30°C ambient temperature, shock mounted for vibration isolation
| Grating | 1080 lines/mm, concave replica on quartz blank, ruled under laser interferometric control and is the only optical surface in the spectrometer | Optics
|
| Wavelength range | 165-800 nm
| Wavelength drive
| computer controlled stepper-motor driven primary slit
| Resolution | Reciprocal linear dispersion is 0.93 nm/mm in 1st order, 0.46 nm/ mm in 2nd order, 0.31 nm/mm in 3rd order, and 0.23 nm in 4th order. Resolution with 20 µm primary and 50 µm secondary slits is 46 pm in 1st order, 23 pm in 2nd order, 16 pm in 3rd order, and 12 pm in 4th order (all FWHM).
| Detector | up to 60 channels, each optical channel consists of a secondory slit, 45 deg. mirror, filter (optional), and photomultiplier tube | Detection mode
| spectral scanning; off-peak background correction
| Read out |
| |
| ICP generator
| compact design integrated into spectrometer housing; solid-state driver; crystal controlled; closed loop autotuning;
| Frequency (MHz) | 27.12
| Power (Watt)
| 750
| Load coil
| four turns, water cooled (recirculating cooling system)
| Torch
| Minitorch with reduced argon consumption (since 1989), fixed one-piece quartz
| Argon flow (L/min)
| < 9 l/min
|
|
| Sample introduction | concentric Meinhard nebulizer; conical spray chamber (45 ml volume for reduced memory effect)
| Operating system and software
| MS-DOS based and written in PASCAL and ASSEMBLER
| Computer
| IBM Personal Syslem/2 Model 30 with 640 Kbyte memory and VGA graphics (since 1989); 20Mbyte fixed disk drive with a 3-1/2 inch diskette; printer is an IBM Proprinter 11, IBM color monitor
| Short term precision (RSD for 10 replicates)
|
| Long term precision (RSD for 4 hours)
|
| Sensitivity
|
| Detection limits (3s, µg/L) |
| Accessories
| Three nebulizer options are offered to greatly improve performance: an efficient, reliable Ultrasonic Nebulizer (USN); a Corrosion Resistant Maximum Dissolved Solids Nebulizer (CRMDSN) designed for high total dissolved and/or suspended solids as weil as for sampies with high concentrations of mineral acids; Hydrofluoric acid; and a Conductive Solid Nebulizer (CSN), which eliminates the need to dissolve metallurgical sampies; | Special features
| Minitorch for reduced argon consumption
|
|
|