Instrument Database:
Claisse, Inc - Peroxide Fluxer
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
Claisse, Inc
|
Categories |
Sample preparation: Fusion |
Prepares: • Solutions for AA, ICP, and wetchemistry analysis
Specially designed for: • Fusing samples in sodium peroxide, sodium carbonate, and potassium pyrosulfate
Processes: • Highly refractory compounds such as chromite, magnetite, ilmenite, rutile, silicon, and siliconcarbide • Materials with high sulfide contents • Noble metals
During the fusion process, the crucibles rock back and forth to mix the contents, thus enhancing heating efficiency and ensuring homogenous melting. When the fusion process is complete, the gas is automatically shut off, the crucibles return to their initial horizontal position and fans start working to speed up the cooling reaction. Crucibles can continue the rocking motion for a few minutes, if wanted. The whole crucible is finally dunked into a suitable acid, which quickly dissolves the produced salts. The solution can now be diluted to the desired volume, and analysis can begin.
Instrument benefits
SUPERIOR ANALYTICAL ACCURACY • Complete dissolution of the samples • Consistent flame and temperature control (100 heating levels) • Uniform heating for reproducible results • Retention of volatile elements • Reduction of acid impurities • Efficient agitation and superior homogenization of the melt • All fusion parameters can be modified: heating & cooling duration, mixing speed & amplitude, gas flow • High-performance at high altitudes
USER-FRIENDLY • Fully automatic • Individual burner selection • Easy loading of crucibles • 7 preset fusion programs available • Product assistance: Claisse chemists are available for assistance with instrument maintenance and method development
ECONOMICAL • Time saving: 5 to 10 minutes to prepare a solution • High throughput: 6 samples at a time with 40 to 60 samples per hour • One-burner-per-sample design • Uses propane or LPG gases only • No compressed air and no oxygen required • Low power consumption and no special outlet required • Allows fusion of a wide array of samples • Operates free of supervision • Low maintenance and repair costs
|
Specifications |
|
Electrical: 100, 115, 230 V; 150W; 50/60 Hz Propane: Input pressure: 0.7 bars (9psi) Maximum consumption per burner: 6 g/min Heat: Maximum per burner: 25,000 BTU/hr Maximum recommended operating condition: 17,000 BTU/hr Dimensions: Main unit: 35 x 63 x 33 cm Controller: 20 x 38 x 12 cm Weight: Approx. 32 Kg (70 lb.)
|
|