Instrument Database:
Gilson, Inc - ASPEC™ XL4 Four-Probe SPE System
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
Gilson, Inc
|
Categories |
Separation: Extraction: SPE |
The ASPEC™ XL4 uses positive pressure elution to automate solid phase extraction for both routine and method development work. The system processes four samples in parallel with the capability of processing up to 50 samples per hour. Its large capacity accesses up to 108 samples in a single batch.
- Accepts 1-, 3- and 6-mL standard SPE cartridges and 3M 96-well microplates
- Transfers extract into open or sealed autoinjector vials
- Provides mixed-mode (sequential, batch or combination) SPE and multicollection capabilities
- Collects eluting fractions into different rows utilizing the multicollect feature
- Uses positive-pressure elution technology, ensuring better reproducibility and higher precision
- Accesses up to nine solvents from any size bottle via transfer ports
|
Specifications |
|
Manufacturing Standards Meets applicable safety and EMC certification standards; UL and CE certified
Communication Interface Eight inputs + one analog signal (10 or 100 mV); four outputs, three relay outputs and 2 x 12 V power supplies
Environmental Conditions Indoor use Altitude: up to 2000 m Temperature range: 5–40°C Air pressure: 75–105 kPa Pollution degree: 1 or 2, in accordance with IEC 66 Humidity: Maximum relative humidity 80% for temperatures up to 31°C, decreasing linearly to 50% relative humidity at 40°C
Flow Rate 1 µL/min to 20 mL/min
Number of Solvents Eight solvent bottles with no volume limit, and 11 rows of four 20 mL tubes on the tray
Power Requirements Frequency: 50–60 Hz Voltage: 100–120 or 220–240V, mains voltage fluctuations not to exceed ±10% of the nominal voltage
Sample Preparation Capability All SPE steps (condition, load, wash, elute) including SPE cartridge drying (gas valve fitted as standard). Multicollect and multiuse methods. Reagent and internal standard addition, derivatization, dilution, mixing. With advanced programming options (link method, sub method).
Sample/Reagent Vessels 3.5-mL (10 x 75 mm), 9-mL (13 x 100 mm) and 20-mL (16 x 150 mm) tubes and 2-mL (12 x 32 mm) autoinjector vials
Software 735 Sampler Software
Syringe Capacity 500 µL; 1, 5, 10 (standard) or 25 mL
Temperature Control Optional Peltier Controller (4° to 40°C)
Transfer of the Extract The extract can be automatically transferred into open or sealed autoinjector vials for further analysis
1-mL SPE Cartridge (100 mg, or 4-mm disk) Sample/cartridge capacity: 108 Max. collection volume: 5 mL Multiple collection: Up to 9 fractions per cartridge
3-mL SPE Cartridge (200–500 mg, or 7-mm disk) Sample/cartridge capacity: 80 Max. collection volume: 15 mL Multiple collection: Up to 5 fractions per cartridge
6-mL SPE Cartridge (500–1000 mg, or 12-mm disk) Sample/cartridge capacity: 28 Max. collection volume: 15 mL Multiple collection: Up to 5 fractions per cartridge
Dimensions (w x d x h) 52 x 66 x 84 cm (21 x 26 x 33 in.)
Weight 18 kg (40 lbs.)
|
|