Knauer GmbH - Smartline PDA Detector 2800 UV/VIS/NIR
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
Knauer GmbH
|
Categories |
Chromatography: LC-Detector: Diode Array |
With a combination source of light for highly sensitive measurements between 190 and 1,020 nm
The Smartline Diode Array Detector 2800 with a UV/VIS/NIR light source supports a wavelength range of 190 to 1,020 nm. Aside of having an enlarged linearity range, this detector is distinguished by having an outstanding signal to noise ratio. Work with spectra libraries is supported by the detector which makes the inspection of peak purity possible. Maintenance and exchange of the measuring cell is made simple by having direct access to it from the front panel.
|
Specifications |
|
Diode Array 1024 diodes in a ceramic chassis
Wavelength Range 190-1,020 nm
Light Source Deuterium and Tungsten-Halogen lamp (combination source of light)
Spectral Resolution 2.4 nm (0.8 nm pixel distance)
Wavelength Accuracy < 0.5 nm
Wavelength Reproducibility < 0.1 nm
Signal-to-Noise Ratio 1 x 10-5 AU, according to ASTM, optical wavelength 10 mm, 1 ml/min methanol, λ= 254 nm/750 nm, Δλ= 4 nm (±2 nm)
Linearity 0-2.0 AU
Drift < 5 x 10-4 AU/h, according to ASTM E1657-94
Technical Parameters Digital Control and Output: Ethernet, Analog output: 2 x ± 10 V, Power Supply: 115/230 V, 50-60 Hz, 100 W
Dimensions / Weight 220 x 185 x 340 mm (W x H x D) / 8.5 kg
|
|