Instrument Database:
Peter Huber Kältemaschinenbau GmbH - minichiller w-2-H1
|
|

|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
Peter Huber Kältemaschinenbau GmbH
|
Categories |
Temperature control: Refrigerated/Heating Circulators |
Chiller with water-cooled refrigerating unit and circulation pump. Evaporator (cooler), tank and housing of stainless steel. Pressure-suction pump made of industrial plastic material. With adjustable overtemperature protection according to DIN 12876.
CC2: The Comfortable Level protection and maximum and minimum setpoint for additional safety, external temperature sensor connection, external temperature control and temperature programmer (1 program with a max. of 5 segments) and easy ramping function. Plug & Play Technology - new generation of microprocessor controlled Compatible Control. Simple operation with a rotary knob and digital display, easy control, clear text, menu-driven, set point limiting, visually and acoustically alarm, mains failure automatic, programmable. Simple user guidance using a 4 line graphical LCD display. Administrator level and multiple user levels are available. All controllers (CC1, CC2, CC3) are backwards compatible, interchangeable and calibrateable. 3 year warrantee (registration necessary).
|
Specifications |
|
Operating temperature range | -20...100 °C |  |  | Resolution of display | 0.1 K |  |  | Temperature stability at -10°C | 1 K |  |  | Temperature adjustment | digital |  |  | Temperature indication | digital |  |  | Internal temperature sensor | Pt100 |  |  | Sensor external connection | Pt100 |  |  | Safety classification | FL |  |  | Heating power | 1 kW |  |  | Cooling power | |  |  | at 15°C | 0.3 kW |  |  | at 0°C | 0.2 kW |  |  | at -10°C | 0.14 kW |  |  | at -20°C | 0.07 kW |  |  | Refrigerant | R134A |  |  | Circulation pump: | Yes |  |  | max. delivery | 12 l/min |  |  | max. delivery pressure | 0.2 bar |  |  | max. suction pressure | 0.1 bar |  |  | Pump connection | M16x1 |  |  | Pump connection for hose | NW8/12 |  |  | Cooling water connection | 1/2" |  |  | Cooling water connection for hose | 3/8" |  |  | Verbrauch bei Wasser 15°C, Vorlauf 15°C | 15 l/h |  |  | min. cooling water differential pressure | 3 bar |  |  | min. filling capacity | 1.4 litres |  |  | Volume of expansion | 2.6 litres |  |  | Overall dimensions WxDxH | 225x360x380 mm |  |  | Net weight | 25 kilogramme |  |  | Power supply requirement | 230V 1~ 50/60Hz |  |  | max. current (1 Phase) | 2 A |  |  | Fuse (1 phase) | 6 A |  |  | min. ambient temperature | 5 °C |  |  | max. ambient temperature | 32 °C |  |  |
|
|