Instrument Database:
Shimadzu Europe - AA-6200
|
|
|
|
|
|
|
Year of introduction |
|
Status |
historical ( out of sale ) |
Company |
Shimadzu Europe
|
Categories |
Spectrometer ( Atom. ): AAS: General |
Shimadzu's new double beam AA-6200 atomic absorption spectrophotometer recombines high performance and affordability. The double beam system provides the superior baseline stability expected from high performance optical systems at an affordable price. Use of WindowsTM 95 and AtomicA's software Wizard function provides easy operation for system control and data processing. The AA-6200 uses the least linear bench space of any atomic absorption spectrophotometer in the world. In spite of its compactness, the performance and capability is superior. In fact, the AA-6200 outperforms other atomic absorption instruments many times its physical size.
|
Specifications |
|
Optics | Double Beam (chopper mirror) | Monochromator | aberration corrected Czerny-Turner monochromator | Holographic grating | (1,600 lines/mm) | Wavelength range | 190-900 nm | Wavelength selection | Automated | Slit | 0.2 nm, 0.7 nm Manual setting | Background Correction | D2-Lamp method | Lamp Turret | 2-lamps simultaneously lit (manual turret) | Lamp Mode | Emission, Non-BGC, BGC-D2 | Nebulizer | integrated impact bead and jacket tip,Pt/Ir capillary | Chamber | Polypropylene | Burner | Fixed back/forward position and burner height, Simply switching of Air/C2H2 and N2O/ C2H2 burner | Burner head
| Titanium 10cm slot burner (Optional high temperature burner) | Gas Control | Manual setting of flow rate, Automatic Air/ N2O switching system | Safety | Gas pressure monitoring to prevent flashback, Automatic flame monitoring, Safety interlock for misuse of burner, Automatic flame extinguish when power failure | Ignition | Push ignite button | Software | Software based on MS-Windows!" | Dimension and weight | W690 x D425 x H370 mm, 38 kg | Power requirements | AC220V, 230V, 50/60Hz, 300 VA, (Requires transformer for AC115V) | Certification
| CE
| Ambient temperature and humidity range | Temperature: 10~35°C Humidity: 45~80% (but less than 70% if temperature is over 30°C
| Standard accessories | Burner Head of 10cm slot. Nebulizer assembly. Deuterium lamp. Drain tube, Teflon tube for sample suction. Hose assembly for Air. Hose assembly for C2H2. Hose band, RS-232C connection cable (9P-9P). AC cable for 220, 230V | Personal Computer requirements |
| Operating environment: | MS-WindowsTM 95 or MS-WindowsTM Ver3.1 | CPU: | 486DX-66 or Higher (In case of MS-WindowsTM3.1), Pentium-75 or Higher (In case of MS-WindowsTM 95) | RAM: | 16MB or higher | CRT: | VGA or Higher | Hard disk: | 120 Mb or larger | 3.5" Floppy disk drive: | One or more units |
|
Related links:
|