Shimadzu Europe - GC-17A Version 3
|
|
|
|
|
|
|
Year of introduction |
|
Status |
historical ( out of sale ) |
Company |
Shimadzu Europe
|
Categories |
Chromatography: GC |
Shimadzu's GC-17A provides the parameter control necessary for the most demanding applications and schedules.
Confidence in Results The GC-17Av.3 ensures that all the operating parameters are set reproducibly, including carrier gas flow, split ratio, all heated zone temperatures, detector combustion and make-up gas flows, and detector output range and current. Analyses are achieved quickly and confidently with electronic flow control of all carrier gas and detector gas functions. Shimadzu's Advanced Flow Control (AFC) of the carrier gas is teamed with Advanced Pressure Control (APC) of the detector gases (up to 14 flow channels), providing the most accurate GC performance available.
Simple Operation Up to ten GC parameter files can be entered, and the GC can be directly controlled, with the intuitive keypad. Parameter data is carefully protected by a password system which limits access to critical settings. Set parameters automatically with the GC clock scheduler, so you can conserve gas at night and have the instrument ready to use in the morning.
This premium GC combines the control and data protection required for ultimate confidence in rapidly obtained results, and is a cost effective solution for laboratories.
Detectors available:
- FID (Flame Ionization Detector)
- TCD (Thermal Conductivity Detector)
- ECD (Electron Capture Detector)
- FTD/NPD (Flame Thermionic Detector, Nitrogen Phosphorous Detector)
- FPD (Flame Photometric Detector)
- PID (Photoionization Detector)
- PDID (Pulse Discharge Ionization Detector)
- ELCD (Electrolytic Conductivity Detector)
|
Specifications |
|
Column Oven
Inner Volume:: 14.4 L (280W x 280H x 184D mm) Temperature Range: Ambient + 4°C - 450°C (1°C increments) -90°C - 450°C with original cryogenic accessories Absolute temperature accuracy: ± 1% of the set temperature (Kelvins) Temperature non-uniformity: Less than 2°C Temperature variation coefficient: ± 0.01°C/°C Double overheat protection: Yes Number of temperature ramps: 5 (cooling ramp possible) Temperature program rate: ~ 100°C/min in 1°C increments Maximum run time: 655 min
Injection port
Number of individually controlled zones: 2 Temperature range: ~ 450°C (1°C increments) Temperature Program: Both heating and cooling step program Double overheat protection: Yes Types available: Split/splitless (standard), direct injection, on-column PTV (programmable temperature vaporizer)
Auxiliary Heated Zones One additional temperature-controlled heated zone (AUX1) standard, with additional 4 zones optional (AUX 2-5).
Detector Three detector zones with individual temperature control Double overhead protection
Flow Control (AFC) Split/Splitless mode: Pressure Range: ~ 400 kPA (in 1 kPA increments); Number of Program Steps: 5; Split Ratio Range: ~ 9,999 (in 1/1 ratio increments)
Direct Injection Mode: Pressure control mode: ~ 400 kPA (in 1 kPA increments); number of program steps: 5; flow rate control mode: ~ 400 mL/min (in 1mL/min increments)
Flow Control (APC) Pressure Range: ~ 400kPA (in 1 kPA increments) Number of program stops: 5
Display Two lines on backlit LCD display Shows temperatures, times, pressure and other parameters.
Dimensions (mm), Weight (kg) 515W x 437H x 520D, 35
Power Requirements 100, 115, 220, 230, or 240 VAC as ordered, 50/60 Hz, 1800 VA
|