The Shimadzu LC-10AS Pump is a solvent delivery unit with a single plunger reciprocating pump, which has been developed for improving accuracy and sensitivity of analysis, for use in a high-performance liquid chromatography.
|
Specifications |
|
- Pump: Single plunger reciprocating pump (Approx. 100ul/stroke)
- Modes of solvent delivery: Constant flow or constant pressure solvent delivery
Constant flow delivery - Range of flow rate: 0.01-5ml/min (10-400kgf/cm2), 5.01-9.99ml/min (10-200kgf/cm2) - Accuracy of flow rate setting: Within ±2% or ±10ul/min whichever is higher (0.1-5ml/min, 10-400kgf/cm2, with water and room temperature at 20°C ) - Precision of flow rate: Within ±03% (0.1-5ml/min, 10-400kgf/cm2 with water and room temperature at 20°C) Constant Pressure delivery - Pressure setting range: 10-400kgf/cm2 (at I kgf/cm2 stepping) - Pressure setting accuracy: Within ±10% or ±10kgf/cm2, whichever is larger - Pressure limit: High/Low pressure
- Dampener: Flow-through type with internal capacity of approx. 1300ul
- Materials in contact with solvent: SUS316, UHMW-PE, Elgiloya, Kalrez, PEEK ETFE, PCTFE, PFFE
- Suction filter: 10um mesh
- Line filter: 2um mesh, capacity approx. 200ul
- Time program: Flow rate, pressure, EVENT, LOOP (repeated programs), 10 files, total 320 steps
- Pressure indication accuracy: Within ±2% or ±5kgf/cm2, whichever is larger
- Plunger cleaning flow line: Cleaning is possible with the connection of syringe and pump
- Size: 260W x 140H x 420D
- Weight: 11kg
- Operating temperature: 5-35 °C
|