Instrument Database:
Shimadzu Europe - TOC-V CSH/CSN/CPH/CPN
|
|
|
|
|
|
|
Year of introduction |
|
Status |
historical ( out of sale ) |
Company |
Shimadzu Europe
|
Categories |
Dedicated Analyzer: Carbon: TC Dedicated Analyzer: Carbon: TIC Dedicated Analyzer: Carbon: TOC |
The TOC-systems TOC-V CSH/CSN and TOC-V CPH/CPN uses the combustion catalytic oxidation and NDIR Method. The 680 °C combustion catalytic oxidation method, that was developed by Shimadzu and subsequently spread around the world, can efficiently analyze all organic compounds. - Integrated sample preparation reduces workload
- Device carry out acidification , dilution and sparging
- Internal dilution function to create multi-point calibration curbes
- Extended working range up to 20000 mg/L
- High sensitivity catalyst for ultra pure water application
- Operation of the TOC-V CSH/CSN stand-alone version via the large integrated control panel display
- Measurement conditions can stored as method in different method groups
- Multi-point calibration up to 10 points
- Results of 200 samples can be stored
- TOC-Control V Software runs the TOC-V CPH/CPN, PC-Version
- Can be upgraded with
- OCT-1 and ASI-V for automation
- TNM-1 for simultaneous TOC and TN measurement
- SSM-5000A for solid samples (suspensions , sludge or swabs)
|
|