Wavelength range
185nm - 3300nm
Spectral Bandwidth
8 steps in ultraviolet/visible region 0.1, 0.2, 0.5, 1, 2, 3, 5, 8 nm
10 steps in near-infrared region 0.2, 0.5, 1, 2, 3, 5, 8, 12, 20, 32 nm
Resolution
0.1 nm
Sampling Pitch
0.01 - 5 nm
Wavelength accuracy
UV/VIS region: ±0.2 nm NIR region: ±0. 8 nm
Wavelength repeatability
UV/VIS region: less than ±0.08 nm NIR region: less than ±0.32 nm
Stray light
< 0.00008% (220 nm, NaI)
< 0.00005% (340 nm, NaNO2)
< 0.0005% (1420 nm, H2O)
< 0.005% (2365 nm, CHCl3)
Photometric range
-6 to 6 Abs
Noise
0.00005 Abs or less (500 nm), 0.00008 Abs or less (900 nm), 0.00003 Abs or less (1500 nm)
Slit width 2nm, RMS value at 1 sec. response
Photometric System
Double beam
Dimensions
1020(W) x 660(D) x 275(H) mm