Instrument Database:
VG Elemental - PQ Eclipse (PQe) ( Thermo Scientific )
|
|
|
|
|
|
|
Year of introduction |
1992 |
Status |
historical ( out of sale ) |
Company |
VG Elemental
|
Categories |
Mass spectrometer: ICP-MS |
Compact low-cost ICP-MS instrument for environmental analysis at reduced sensitivity as a replacement for ICP-AES and ETAAS
|
Specifications |
|
Dimensions (w x d x h in cm)
| 138.5 x 70 x 107.5
| Weight (kg)
| 350
| Power
| 208-240 V, single phase, 35 A | Mass filter
| Quadrupole
| Mass range
|
| Resolution
|
| Detector | Faraday cup, analogue mode | Dynamic range (orders of magnitude) | 7 (+- 10%) | Vacuum system
| turbomolecular pump | ICP generator | free-running | Frequency (MHz)
| 27.12 | Power (Watt)
| 1800 | Load coil
|
| Torch
|
| Argon flow (L/min)
|
| Sample introduction |
| Operating system and software
| OS/2Software provides on-line help, build-in quality control routines, overnight operation | Computer |
| Short term precision (RSD for 10 replicates)
| < 1% for Be,Co,In,Pb,U at 1 ppm | Long term precision (RSD for 4 hours) | < 4% for 4 hours, < 5% for 24 h (most elements) | Sensitivity (Mcps/ppm)
|
| Detection limits (3s, ng/L)
| 200 for U, 300 for Be,Co,In,Bi; 500 for Pb | Isotope ratio precision (RSD for Ag-107/Ag-109)
|
| Doubly charged Ba ions |
| Ce Oxide ions
|
| Accessories
| 220 samples autosampler | Special features
| solutions with a dissolved solids content of up to 0.5% can be analysed; an "all element" acquisition, for more than 75 elements can be carried out in < 90 seconds |
|