The Waters 2790 Alliance is an integrated solvent and sample management platform which combines two traditional high-performance liquid chromatography (HPLC) components: a solvent management system and a sample management system
Dry Prime/Wet Prime: Automatic, front panel control
Flow Ramping: Time (0.01 to 30.00 min in 0.01-min increments) to reach maximum flow rate
Maximum Operating Pressure: 5000 psi (345 bar) at 0.010 to 3.000 mL/min; Programmable upper and lower limits
Composition Range: 0.0 to 100.0% in 0.1% increments
Composition Accuracy: ±0.5% absolute, independent of backpressure (Proportioning Valve Pair Test [degassed methanol: methanol/propylparaben, 2 mL/min, 257 nm])
A Composition Precision: <=0.15% RSD o r<=0.02 min SD, whichever is greater, based on retention time. (Degassed 60% methanol:40% water Dial-a-Mix, 1 mL/min, six replicates, phenone mix, 257 nm)
Flow Precision: <=0.075% RSD or <=0.02 min SD, whichever is greater, based on retention time (N = 6) or volumetric measures (0.200 to 5.000 mL/min), isocratic premix
Flow Accuracy: <=1% or 10 µL/min, whichever is greater, (0.200 to 5.000 mL/ min), degassed methanol, at 600 psi backpressure
Sample Management System Specifications
Number of Sample Plates: Total of four plates: - Microtiter plate – 96- or 384-well - Vial plate – 2.0-mL vial (48) - Tube plate – 0.5-mL microcentrifuge tube (48) or 1.5-mL microcentrifuge tube (24) - Open access plate – 2.0-mL vial (24 or 48)
Sample Temperature Control: 4 to 40° C, programmable in 1° C increments
Maximum Sample Capacity: 1536 in four 384-well plates
Number of Sample Injections: 1 to 99 injections per sample
Sample Delivery Precision: <0.3% RSD, full loop 50 µL. (Default wash/purge conditions, degassed 60% methanol: 40% water Dial-a-Mix, 1 mL/min, six replicates, paraben mix, 257 nm)
Sample Volume Linearity: >0.999% correlation, 5 to 20 µL, partial fill in 50-µL loop, 3-µL pre- and post-sample air gaps (Default wash/purge conditions, degassed 60% methanol:40% water Dial-a-Mix, 1 mL/min, six replicates, paraben mix, 257 nm)
Needle Wash Solvents: Two: wash (strong solvent) and purge (sample- compatible) solvent A
Sample Carryover: <0.01% or <2.5 nL (default wash/purge volumes), whichever is greater. (Blank injection, 25 µL, 60% methanol:40% water, after six replicates, 25 µL propylparaben, 1 g/L, 257 nm)