Nippon Instruments Corporation - MA-2000 Fully-automatic heat-vaporization mercury measuring equipment
|
|
|
|
|
|
|
Year of introduction |
|
Status |
historical ( out of sale ) |
Company |
Nippon Instruments Corporation
|
Categories |
Dedicated Analyzer: Mercury |
Fully-automatic heat-vaporization mercury measuring equipment The model MA-2000 is intended for heat-vaporizing mercury present in a sample and then collecting it as a gold amalgam for measurement. This model does not require any wet pretreatment of samples, and hence can easily measure solid and liquid samples in as short a time as five minutes per sample. Both the control and data processing software are compatible with Windows, and labor-saving is achieved with the automatic sample changer.
|
Specifications |
|
Heat-vaporization mercury measuring equipment (MA-2)
Pretreatment unit Sample: Solid, liquid, gaseous (Collecting tube used) Carrier gas: Cleaned air Combustion tube: Ceramic Sample boat: Ceramic (10 supplied) Heating temperature: 900°C or above
Detection unit Measurement principle: Non-dispersive double-beam cold-vapor atomic absorption Wavelength: 253.7 nm Lower detection limit: 0.002 ng Measuring range: Max. 1000 ng
Control and data processing system (PC-Win/M-A2)
OS: Windows2000/XP Display: Peak wave form, calibration curve graph, measurement results, measuring status Control: Decomposition: Heating temperature, time Measurement: Measuring object, start/stop, time Processing: Calibration curve, unit setting, concentration calculation, baseline correction, statistical calculations Printable items: Memo, calibration diagram and expressions, measurement table, statistical calculations, peak wave form Accuracy control: Standard check, blank check
Options For heat-vaporization measurement Automatic sample changer (BC-1) Sample Q’ty: 32 & 40 samples Sample carry-in unit: Measure against external pollution (maintains positive pressure)
For reducing-vaporization measurement Reducing-vaporization unit (S-MA) Sample container: 20 mL
Reagent dispensing burette (RD-3) Dispensed liquid: Sulfuric acid, stannous chloride solution, distilled water
Automatic sample changer (SC-3) Sample Q’ty: 60 positions Sample container: 5 mL
For atmospheric mercury measurement Mercury collecting tube bake out oven unit (RH-MA) Heating method: Anti-pollution heating method
|
|