Instrument Database:
Netzsch-Gerätebau GmbH - DIL 402 PC Dilatometer
|
|
|
|
|
|
|
Year of introduction |
|
Status |
historical ( out of sale ) |
Company |
Netzsch-Gerätebau GmbH
|
Categories |
Thermal Analysis: DMA |
The newly developed DIL 402 PC dilatometer combines easy operation, high adaptability to different applications and outstanding performance in one instrument. The optimized design of the measurement system with inductive transducer offers a high degree of reproducibility even without an external chiller.
The horizontal design with easy to move furnace makes it simple to place samples into the large recess of the tube-type sample carrier, even with less than ideal sample geometries. A thermocouple in direct proximity to the sample yields reproducible temperature measurement. This also allows use of the c-DTA® Software for calculation of endothermic and exothermic effects in the sample as well as determination of all the characteristic expansion values.
The interchangeable furnaces with temperature ranges up to 1200°C and 1600°C serve varied applications for expansion control in solids in the areas of quality and production control.
|
Specifications |
|
- Temperature range: RT ... 1200°C, 1600°C (2 furnace types)
- Heating and cooling rates: 0.01 K/min ... 50 K/min (dependent on furnace)
- Sample holder:
fused silica <1100°C Al2O3 <1600°C
- Measuring range: 500/5000 µm
- Sample length: max. 50 mm
- Sample diameter: max. 12 mm
- Δl resolution: 8 nm
- Atmospheres: oxidizing: static and dynamic, inert: dynamic
- Gas flow controller for purge gas (optional)
- c-DTA® for the calculated DTA-signal from dilatometer measurements, ideal for temperature calibration (optional)
- Integrated electronics
- USB interface for PC connection
|
|