Hitachi - Science & Technology - LaChrom ULTRA L-2485U Fluourescence Detector
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
Hitachi - Science & Technology
|
Categories |
Chromatography: LC-Detector: Fluorescence |
For unique sensitivity in the trace range The L-2485U Fluorescence Detector has a uniquely high signal-to-noise ratio even at the fastest detector response times. Additional functions such as variable slit width and adjustable photomultiplier voltage enable further optimisations of detector sensitivity. The L-2485U Fluorescence Detector is equipped with a mercury source for automatic wavelength monitoring and calibration of both excitation and emission wavelengths. For ease of maintenance, all lamps and the flow-through cell are accessible from the front. - Highest possible sensitivity
- Very low drift
- Uniquely low response times
- Extremely high data collection rate
- Minimal internal volumes
- Recording of fluorescence spectra
- Mercury lamp for wavelength calibration
- Lamps and flow-through cell accessible from the front
|
Specifications |
|
Source
| Xenon-lamp (150 W), Hg lamp for wavelength calibration
| Wavelength range
| Excitationb: 200-850 nm, emission: 250-900 nm (at wavelengths > 731 nm another photomultiplier has to be used)
| Spectral bandwidth
| Excitation: 15 nm Emission: 15 nm or 30 nm, adjustable
| Sensitivity
| S/N for the Raman peak of water: > 550 (baseline method, emission slit width 30 nm) for vthe high-speed cell, > 900 for the standard cell (option)
| Response time constants
| 0.01/0.02/0.05/0.1/0.5/1/2 s
| Data collection rates
| Shortest data interval 10 ms or 100 Hz
| Programmes
| Programmable parameters: - Excitation nwavelength
- Emission wavelength
- Baseline processing (auto-zero, hold)
- Photomultiplier amplification
| Flow-through cell
| 3 µL (ultra-high-speed cell)
| FDA/GXP functions
| - Automatic wavelength calibration with various Hg spectral lines
- Maintenance logbook: date of last lamp replacement (Xe and Hg lamps), lamp energy, lamp usage, number of ignitions
In conjunction with the EZChrom Elite software: - Documentation of instrument serial number and configuration
- Detaled audit trails and logbooks
| Safety
| Leak sensor, error handling
| Power requirements
| 1oo-240 V (U50/60 Hz), 260 W
| Dimensions
| 340 (W) x 400 (D) x 300 (H) mm
|
|
|