Instrument Database:
Jobin Yvon Horiba - ULTIMA ( HORIBA Jobin Yvon )
|
|
|
|
|
|
|
Year of introduction |
1997 |
Status |
historical ( out of service ) |
Company |
Jobin Yvon Horiba
|
Categories |
Spectrometer ( Atom. ): AES: ICP-AES |
Specifications |
|
Dimensions (w x d x h in cm)
| 180 x 72 x 59 | Weight (kg)
| 275
| Power
| Single phase, 3 kW, 30 A, 220/240 V, 50/60 Hz
| Spectrometer
| 1-m Czerny-Turner monochromator, nitrogen purged, thermostated
| Grating | ion-etched holographic master, 110 x 110 mm, 2400 lines/mm used in first/second order | Optics
| | Wavelength range | 120 - 800 nm | Wavelength drive | direct drive grating mount; Full range in 10 s; reproducibility +- 0.4 pm
| Resolution | Slit widths of the monochromator are computer selected from the possible choices of 10 or 20 µm entrance and 15 or 80 µm exit widths, as is the choice of HDD and order sorting filters.
| Detector | High Dynamic Detector (HDD)
| Detection mode
| | Read out | |
| | ICP generator
| solid state generator, frequency stabilized; reflected power control | Frequency (MHz) | 40.68
| Power (Watt)
| 2600 | Load coil
| water cooled
| Torch
| fully demountable, radially viewed
| Argon flow (L/min)
| 12 (7 low-flow option)
|
| | Sample introduction | thermoregulated sample compartment; cassette design for easy exchange of the torch
| Operating system and software
| DOS 6.22, Windows 95 compatible software
| Computer
| | Short term precision (RSD for 10 replicates)
| | Long term precision (RSD for 4 hours)
| | Sensitivity
| | Detection limits (3s, µg/L) | | Accessories
| hydride generator; HF-resistant nebulizer and injector, ultrasonic nebulizer
| Special features
| |
|
Related links:
|