Instrument Database:
Australian Scientific Instruments Pty. Ltd. - RESOlution
|
|
|
|
|
|
|
Year of introduction |
2008 |
Status |
available |
Company |
Australian Scientific Instruments Pty. Ltd.
|
Categories |
Sample introduction: Laser ablation |
Exceptional Laser Ablation Laser Ablation - Inductively Coupled Plasma - Mass Spectrometry (LA-ICP-MS) is an analytical technique for determining the chemical and isotopic composition of solid samples, especially prominent in the fields of geology and petrology, but also materials science and increasingly, due to its imaging capability, the life sciences. The RESOlution laser ablation systems are the only LA-ICP-MS platform to offer the industry leading Laurin Technic sample cells and pioneering Geostar µGISTM software for exceptional ablation. They are in use with all mass spectrometers supplied by leading ICP-MS manufacturers. Features:
- High Quality & Robust Design
- Ultra-fast washout times
- Outstanding signal sensitivity
- Highly uniform signal across the entire cell
- Minimal fractionation
- Completely automated instrument control
- Integrated on-sample fluence calibration and laser energy management
- Visual-Stage-Correction offers sub-micron positioning
- Import images from any source, such as an SEM or petrographic microscope
|
Specifications |
|
RESOlution Model
| -HR
| -LR
| -SE
| Laser Safety
| Compliant to FDA/CDRH 21 CFR, Class I laser system, laser fully interlocked, fully enclosed beam path
| Laser Wavelength
| 193 nm
| Laser Pulse Width
| 20 ns
| 20ns | 5-7 ns | Internal Energy Meter
| Yes (closed-loop feedback)
| Maximum Pulse Energy
| 200 mJ
| 200 mJ
| 10 mJ
| Maximum Average Power
| 12 W
| 4 W
| 1.5 W
| Pulse-to-Pulse Energy Stability
| < 1% RSD
| < 1% RSD
| < 2% RSD
| Maximum Repetition Rate
| 100 Hz
| 20 Hz
| 300 Hz
| Cooling
| Water (2-3 L/min @15-20°C) | Air | Air | Maximum Fluence
| 30 J/cm2
| 20 J/cm2
| Round/Square Spot Sizes
| 4-380 µm 4-380 µm
| 2-100 µm
| Rectangular Slit Sizes
| Length: 4-380 µm Width: 4-380 µm
| Length: 2-100 µm Width: 2-100 µm
| Stage Movement Range
| M50A: 50 x 50 mm; S155: 155 x 105 mm
| Stage Reproducibility
| < 2 µm (2 sigma)
| System Weight
| 1240 kg
| 725 kg
| System Width
| 179 cm
| 119 cm
| System Depth
| 89 cm
| 89 cm
| System Height
| 195 cm
| 172 cm
| | | | |
|
|