Beijing Rayleigh Analytical Instrument Corporation (BRAIC) - WFX-110A/120A/130A Atomic Absorption Spectrophotometer
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
Beijing Rayleigh Analytical Instrument Corporation (BRAIC)
|
Categories |
Spectrometer ( Atom. ): AAS: Flame/Furnace |
FEATURES
- The patented flame analysis technique adopting oxygen-rich air-acetylene flame as the substitution for nitrous oxide-acetylene flame for high temperature element analyses. (Only for WFX-110A)
- Automatic hollow cathode lamp changeover. The high performance HCL can be used directly.
- Auto-changeover and position optimization of integrated flame/graphite furnace atomizer.
- Auto-ignition, automatic spectral bandwidth selection. · Fully automated wavelength scanning and peaking
- Light-controlled graphite furnace power supplier, adopting FUZZY-PID temperature control technique and dual curve working mode, ensures fast heating, accurate and stable temperature control and good temperature reproducibility. Temperature auto-correction function provided.
- Graphite furnace with pneumatic control and pressure lock ensures constant pressure and reliable contact.
- Perfect safety protection measures, alarm and automatic safety protection to fuel gas leakage, air pressure deficiency and abnormal flame extinction; spray chamber made of explosion-proof material, all-titanium burner and explosion-proof stopper for flame analyses; alarm and protection function to argon pressure deficiency, insufficient cooling water supply, over-heating and over-current etc. in graphite furnace analyses.
- Adopting large-scale programmable logic array and Inter IC bus technique and highly reliable adapters for electric design.
- Easy and flexible WINDOWS operating system, with multi-windows processing technology to observe the signal profile, the measured results and the calibration curve at the same time. Various analytical programs, data, signals and graphics can all be stored for later use.
|
Specifications |
|
Main Specification Wavelength range 190-900nm
Wavelength accuracy ±0.25nm
Resolution Two spectral lines of Mn at 279.5nm and 279.8nm can be separated with the spectral bandwidth of 0.2nm and valley-peak energy ratio less than 30%.
Baseline stability ≤0.004A/30min
Background correction The D2 lamp background correction capability at 1A is better than 30 times. The S-H background correction capability at 1.8A is better than 30 times. (only for WFX-110A/120A)
Hollow Cathode Lamps Lamp turret 6-lamp turret (WFX-110/120), 4-lamp turret (WFX-130) auto alignment, fully automated scan and peak-picking.
Lamp current adjustment Automatic adjustment and display. Wide pulse current: 0~25mA, Narrow pulse current: 0~10mA Lamp power supply mode Powered by two pulses simultaneously
Optical System Monochomator Single beam, Czerny-Turner design grating monochromator
Grating 1800 l/mm
Focal length 277mm
Blazed Wavelength 250nm
Spectral Bandwidth 0.1nm, 0.2nm, 0.4nm, 1.2nm automatic change.
Atomizer Flame/Graphite furnaceFlame/Graphite furnace atomizer auto changeover
Detection and Data Processing System Detector Photomultiplier with high sensitivity and wide spectral range.
Software Ms-Windows 98/Me operating system
Analytical method Working Curve Method (1st order and 2nd order fitting), Sensitivity auto-correction, Standard Addition Method.
Concentration Auto-calculation of concentration and content
Repeatable measurement Maximum 20 times of repeat measurement, automatic calculation of mean value, standard deviation and relative standard deviation.
Multi-task function Sequential measurement of multi-elements in one sample
Condition reading With model function
Result printing Measurement data and final analytical report printout, editing with Excel.
Flame Analysis Burner 10cm single slot all-titanium burner
Spray chamber All-plastic spray chamber.
Nebulizer High efficiency glass nebulizer with metal sleeve.
Position adjustment Adjusting mechanism for vertical, horizontal positions and the rotation angle of the burner.
Protection for gas system Fuel gas leak protection, abnormal flame extinction alarm, air-pressure deficiency protection and explosion-proof device.
Flame ignition Automatic
Characteristic Concentration and Detection Limit Normal Air-C2H2 flame Cu: Characteristic concentration≤0.025mg/L, Detection limit≤0.006mg/L; Oxygen-rich Air-C2H2 flame Ba: Characteristic concentration ≤ 0.22mg/L Al: Characteristic concentration ≤ 0.4mg/L (for WFX-110A)
Hydride generation Hg: detection limit≤0.5ug/L
Graphite Furnace Analysis Temperature range 20~3000ºC
Light-controlled temperature range 1500ºC -3000ºC, temperature auto-correction
Temperature control accuracy ≤1%
Heating rate 2000°C/s
Characteristic mass Cd≤0.5 ×10-12g, Cu≤5 ×10-12g, Mo≤1×10-11g
Precision RSD≤4%
Safety protection Protections for over-heating, over-current, and less cooling water and argon pressure monitoring functions
Power Supply 220V, 200W
Dimensions 1020 (L) × 492 (W) × 537 (H) mm (main unit) 420 (L) × 420 (W) × 460 (H) mm (graphite furnace)
Weight Unpacked 80kg (main unit), 50kg (graphite furnace)
|
|