nanoLC-Ultra 1D system specifications Dimensions
14” x 24” x 18” (W x D x H)
Weight
70 lbs.
Working temperature
15º to 30ºC
Power
100-240 V AC, 400 VA
Maximum pressure
10,000 psi
Standard flow rate ranges
- Loading pump: 1-30 µl/min
- Nano gradient: 50-500 nl/min (up to 1000 nl/min at reduced maximum pressure)
- Micro gradient: 1-10 µl/min (up to 20 ul/min at reduced maximum pressure)
Flow rate accuracy
<1%
Flow rate precision
< 0.5% @ 200 nl/min
Gradient accuracy
<1% @ 200 nl/min
Gradient delay
<25 nl
Column oven
ambient + 5ºC to 40ºC (±1 ºC)
Built-in 10-port valve
- 1/32” connection; 0.1 mm bore; port-to-port
- Volume < 25 nl; max. pressure 10,000 psi
- Stainless Steel with proprietary coating on valve stator
Wetted parts
Valve stator
Stainless Steel, Titanium, PEEK, Fused Silica, FEP, proprietary inert coating
I/O
1 x USB for instrument control
2 x RS232 serial output
Each pump channel
AD converter Instrument control
Eksigent software with plug-ins for Thermo Electron’s Xcalibur®, Bruker Daltonics’ HyStar® and ABI/Sciex Analyst® software packages
AS-2 high pressure autosampler specifications Dimensions
14” x 24” x 14” (W x D x H)
Weight
up to 70 lbs., depending on configuration
Working temperature
15º to 30ºC
Power
100-240 VAC; 200 VA
Capacity
2 microtiter plates (96 high/low or 384 low) or 48/12 sample vial racks
Injection valve
- 1/32“ connection; 0.1 mm bore; port-to-port
- Port-to-port volume < 25
- Max. pressure 10,000 psi
- Stainless steel with proprietary coating on valve stator and rotor
Injection volume
programmable from 100 nl to 10 µl with standard 10 µl loop. Larger injection volumes possible with optional loop.
Dispenser syringe 25 µl standard; optional : 100 µl, 250 µl
Injection volume reproducibility
- full loop: RSD <0.8%
- partial loop: RSD <1%
- µl-pick-up: RSD <2%
Carry-over
<0.05%
Sample tray cooling
4ºC to ambient -3ºC (±2 ºC)
Wetted parts
Stainless Steel, PTFE, PEEK, Glass, Teflon, TEFZEL, fused silica, proprietary inert coating
I/O
- RS232 serial port for instrument control
- 1 programmable relay output
- 2 programmable TTL inputs
Instrument control
Eksigent software with plug-ins for Thermo Electron’s Xcalibur®, Bruker Daltonics’ HyStar® and ABI/Sciex Analyst® software packages