Instrument Database:
Jobin Yvon Horiba - ULTIMA 2 ( HORIBA Jobin Yvon )
|
|
|
|
|
|
|
Year of introduction |
2001 |
Status |
historical ( out of sale ) |
Company |
Jobin Yvon Horiba
|
Categories |
Spectrometer ( Atom. ): AES: ICP-AES |
Productivity & Performance Defined - Single view radial provides lowest detection limits of any ICP OES spectrometer available today.
- Continuous wavelength coverage from 160 - 800nm standard with optional extension down to 120nm for Far UV analysis of the halogen elements or the use of alternate wavelengths.
- Unmatched spectral resolution of 5 pm at <320 nm minimizes interferences, with optional grating to further improve resolution for line rich matrices such as geological or precious metals.
- Fast, accurate analysis with Rapid Chip electronics and powerful Analyst™ software.
- Patented HDD® detection provides analysis of sub-ppb to percent level in one analysis.
- Dynamic range of up to 10 orders of magnitude using Win-IMAGE.
- Quick-release torch is fully demountable and needs no alignment.
- Unique sheath gas feature, allows continuous analysis of 30% dissolved solids.
|
Specifications |
|
Dimensions (w x d x h in cm)
| 180 x 72 x 59 | Weight (kg)
| 275
| Power
| Single phase, 3 kW, 30 A, 220/240 V, 50/60 Hz
| Spectrometer
| 1-m Czerny-Turner monochromator, nitrogen purged, thermostated
| Grating | ion-etched holographic master, 110 x 110 mm, 2400 lines/mm used in first/second order | Optics
| | Wavelength range | 120 - 800 nm | Wavelength drive | direct drive grating mount; Full range in 10 s; reproducibility +- 0.4 pm
| Resolution | 10/5 pm in first/second order
| Detector | High Dynamic Detector (HDD)
| Detection mode
| | Read out | |
| | ICP generator
| solid state generator, frequency stabilized; reflected power control, air cooled | Frequency (MHz) | 40.68
| Power (Watt)
| 2600 | Load coil
| water cooled
| Torch
| fully demountable, radially viewed
| Argon flow (L/min)
| 12 (7 low-flow option)
|
| | Sample introduction | thermoregulated sample compartment; cassette design for easy exchange of the torch, sheat gas for reduced memory effects
| Operating system and software
| Windows 95 compatible software
| Computer
| | Short term precision (RSD for 10 replicates)
| | Long term precision (RSD for 4 hours)
| | Sensitivity
| | Detection limits (3s, µg/L) | | Accessories
| hydride generator; HF-resistant nebulizer and injector, ultrasonic nebulizer
| Special features
| |
|
Related links:
|