Instrument Database:
Sedico Ltd. - Phoenix-2200DPCV UV/Vis Spectrophotometer
|
|
|
|
|
|
|
Year of introduction |
|
Status |
available |
Company |
Sedico Ltd.
|
Categories |
Spectrometer ( Molec. ): UV/VIS |
The Phoenix-2200DPCV is the top of the range of our UV/Visible spectrophotometers. It is a Double-Beam Ratio Recording instrument with high performance and resolution thanks to its advanced optical system. It is a computer controlled spectrophotometer which comes with its dedicated UV-WIN Windows-based software for commanding the fuctions of the instrument and its accessories from a PC and for data collection and management. We have recently introduced an identical model, the Phoenix-2200DPC which has exactly all the same features as the above with a minor variation in the specifications whereas the bandpass is a fixed 2nm instead of being variable in 6 values down to 0.1nm. The price of the Phoenix-2200DPC is sensitively lower than the Phoenix-2200DPCV. . A wide and comprehensive range of accessories for routine and/or special applications is available upon demand. For further information, please apply to your nearest distributor or call our products specialists.
|
Specifications |
|
SPECIFICATIONS Hardware: - Wavelength Range : 190-900 nm
- Spectral Bandwidth: Variable for the Phoenix-2200DPCV at 0.1, 0.2, 0.5, 1.0, 2.0, 5.0nm, Fixed for the Phoenix-2200DPC at 2nm.
- Wavelength Accuracy: +0.3nm(automatic wavelength correction)
- Wavelength-Reproducibility: 0.1nm
- Photometric Accuracy: +0.002Abs(0-0.5Abs), +0.004Abs(0.5- 1.0Abs), +0.3%T(0-100%T)
- Photometric-Reproducibility: +0.001Abs
- Stray Light : 0.02%(at 220nm 10g/l NaI aq.)
- Baseline Flatness: +0.001Abs
- Noise : < 0.0004Abs (at 500nm,with bandwidth 2nm)
- Stability : 0.0004Abs (2 hours after power on)
- Photometric Mode : Absorbance, Transmission, Concentration
- Photometric Range : -4.0 - +4.0Abs
- Lamp-Interchange: Automatically change at user's preset wavelength
- Monochromator : High Quality Blazed Holographic grating.1200L/mm
- Detector: Photo Multiplier R928-40
- Sample Compartment : 130x240x120mm, distance between light beams 100mm
- Light Source : Socket halogen lamp and deuterium lamp(instrument does not need any manual adjustments after changing lamp)
Software: - Basic Measurement Mode: Photometric, Spectrum, Quantitation, Kinetics.#
- Photometric : Photometic measurement with result at single-wavelength and multi-wavelength of 10 wavelengths Max.
- Quantitation : Quantitation of single-wavelength, two- wavelengths, three-wavelengths and derivative with multi- Point calibration curve ,1st to 4th order calibration curve regression.
- Spectrum : Spectrum scan and arithmetic calculation, smoothing, 1st to 4th derivative,peakpick,point readout,range expanding.
- Kinetics : Time scan of 10 wavelengths max. and arithmetic calculation,smoothing,1st to 4th derivative, peakpick, point readout, range expanding.
- File-Function: Load and save of data file and parameter file
|
|